Beurer · BM

BM 64

Release date not published

An upper-arm blood pressure monitor with AFib screening, two-user memory, triple-measurement mode and Bluetooth transfer to a phone.

Type
Upper arm
Users
2
App
Yes
Validated
–

What reviewers say

GadgetGuy liked automatic phone sync and useful long-term records, though operation needed some care. Computer Bild praised fast measurements and triple-measurement mode, but found blood-pressure and pulse readings less precise than the best devices in its test, and criticised the weight and lack of guest mode.

From gadgetguy.com.au and computerbild.de

Measuring

Type
Upper arm
Irregular heartbeat detection
Yes
AFib detection or ECG
Yes
Method
Automatic oscillometric upper-arm measurement during inflation. UK item 65400, EAN 4211125654009; manual type BM 64. Announced on 12 November 2023 with planned German availability from March 2024; actual first-sale date unresolved
Accuracy
Manual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse. Systolic range 50–280 mmHg, diastolic 30–200 mmHg and pulse 40–199 beats/minute; maximum cuff pressure 300 mmHg is separate
Clinical validation
Current manual claims clinical testing to ISO 81060-2; independent validation for this exact model not established. The shared launch release's pregnancy-validation claim belongs to BM 53, not BM 64
Features
Advanced arrhythmia detection flags possible atrial fibrillation, without recording an ECG. Coloured classification scale, cuff-position check, HSD resting indicator and low-battery indication. Maker retains triple-measurement mode; May 2024 instructions describe three measurements with 30-second waits and a final average, while September 2026 instructions omit that operation section

Cuff

Cuff fits
22–42 cm
Cuff
Supplied premium cuff fits upper-arm circumferences of 22–42 cm; conventional tube connection, not easyClip or easyLock

Connections and memory

Users
2
Readings stored
120 per user
Bluetooth and app
Yes
Memory
Two separately selected users with 120 readings each; oldest reading replaced when full. Recall individual readings and average of all stored readings, plus morning (05:00–09:00) and evening (18:00–20:00) averages over the last seven days. Date/time and user-selection slider explicitly described in the manual
App
Bluetooth transfer to beurer HealthManager Pro with automatic transfer configurable on the monitor; six-digit pairing PIN. Manual publishes 2400–2483 MHz band and maximum 5 dBm transmission power, without a Bluetooth version
Power
Four included 1.5 V LR6 AA batteries; approximately 300 measurements depending on blood pressure, inflation pressure and Bluetooth connections. Optional USB-C mains supply 5 V DC / 1 A; remove batteries for mains use. Included USB-C cable does not charge batteries

Design

Display details
Illuminated inverse XL LCD with white digits on black; four sensor touch buttons and a central START/STOP control. Manual IP20
Size
143 × 105 × 60 mm (L × W × H). Manual and datasheet approximately 450 g with cuff and without batteries; current UK website lists 646 g excluding packaging without specifying cuff/battery weighing conditions
In the box
Monitor, 22–42 cm premium cuff, four AA batteries, USB-C cable, storage pouch and instructions; mains adapter optional

Pros and cons

Pros

  • AFib screening and triple-measurement mode
  • Two users with 120 readings each
  • Illuminated display and Bluetooth transfer

Cons

  • No ECG recording
  • Mains adapter sold separately

Closest rivals

Model
TypeUsersAppValidatedReleased
Upper arm1YesYes–
Upper arm2Yes––
Upper arm2Yes––

Sources

  1. beurer.com/uk/p/65400
  2. res.cloudinary.com/beurer/image/upload/v1789966771/stibo-live/65400_BM64_2026-0…
  3. res.cloudinary.com/beurer/image/upload/v1744513867/stibo-live/65400_BM64_2024-0…
  4. res.cloudinary.com/beurer/image/upload/v1744597369/stibo-live/65400_BM64_2023-1…
  5. res.cloudinary.com/beurer/image/upload/v1719377596/stibo-live/65400_BM64_2023-0…
  6. beurer.com/de/presse/2023/Beurer-Medica-Blutdruck.php

Photo: Official manufacturer photograph; complete supplied canvas and background preserved.