A&D Medical UA-611Plus vs Microlife BP A150 AFIB
Better figure, where one is clearly better
| A&D MedicalUA-611Plus | MicrolifeBP A150 AFIB | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Upper arm | Upper arm |
| Clinically validated | Yes | Yes |
| Irregular heartbeat detection | Yes | Yes |
| AFib detection or ECG | Yes | Yes |
| Method | Automatic oscillometric measurement with fuzzy-logic-controlled inflation; designed for adult home use | Oscillometric upper-arm measurement during deflation, with Gentle+ pressure control. Side switch selects standard single measurement or AFIB/MAM mode; three automatic measurements with 15-second pauses are analysed into one displayed result. Individual readings are not displayed; a fourth measurement is taken if one is questionable. First actual on-sale date unresolved |
| Accuracy | Maker specification ±3 mmHg pressure and ±5% pulse; pressure range 0–299 mmHg, systolic 60–279 mmHg, diastolic 40–200 mmHg and pulse 40–180 beats/min | Exact S-V9 manual gives blood pressure 20–280 mmHg, pulse 40–200 bpm, static pressure ±3 mmHg and pulse ±5%. Current global support instead separates systolic 60–255 and diastolic 40–200 mmHg. Slovenian page's 20–80 blood-pressure line conflicts with both and is not used. Cuff-pressure display 0–299 is separate |
| Clinical validation | STRIDE BP lists UA-611Plus as equivalent to the validated UA-1020, citing 2013 BHS 1993 studies; this is approved device equivalence, not a direct UA-611Plus validation study. A&D's shared manual separately states ISO 81060-2:2013 clinical testing | STRIDE BP lists equivalence to BP A100, adult and diabetes populations, and 22–42 and 32–52 cm cuffs. Headline reflects registry acceptance by equivalence, not an exact-model validation study. Broad maker special-population and AFib performance claims are not treated as independently established exact-model trials |
| Features | IHB/AFib screening indicator, pressure progress bar, WHO classification indicator, stored-reading average and automatic power-off; maker says AFib screening is for monitoring and not diagnosis | AFib indication is active only in AFIB/MAM mode, not the standard single-measurement mode. It is a screening indication rather than ECG recording or diagnosis. Cuff-fit and arm-movement indicators, classification scale, low-battery warning and automatic switch-off after approximately one minute |
| Cuff | ||
| Cuff fits | 22–32 cm | 22–42 cm |
| Cuff | Supplied latex-free and metal-free cuff fits 22–32 cm; optional small 16–24 cm, semi-large 23–37 cm, wide-range 22–42 cm and large 31–45 cm cuffs, sold separately | Slovenian maker bundle supplies conical M-L 22–42 cm cuff. Manual lists optional S 17–22, M 22–32, L 32–42 and L-XL 32–52 cm sizes and optional preformed cuffs; these are not all supplied |
| Connections and memory | ||
| Readings stored | 60 (better) | 1 |
| Bluetooth and app | No | No |
| Memory | Last 60 blood pressure and pulse measurements, with stored-reading average; no separate user memories listed | Original BP A150 AFIB S-V9 guide stores the last result only; recall by holding ON/OFF while switched off until the stored result and M appear. Slovenian product table instead lists 30 places, and global support's other manual filename refers to A150-30 AFIB. The separately named A150-30 model's history is not transferred to this original record. No clock or user-selection operation established |
| App | No Bluetooth; readings can be recorded manually | – |
| Power | Four AA 1.5 V batteries according to the exact manual; approximately 700 measurements with LR6 alkaline or 200 with R6P manganese batteries at 180 mmHg and 23°C. UA-611Plus has no DC socket; the adapter in the shared manual applies only to UA-651Plus | Four 1.5 V AA alkaline cells; approximately 920 measurements with new batteries in S-V9 guide, with no separate AFIB/MAM-cycle count stated. Optional DC 6 V, 600 mA Microlife adapter stops battery-current consumption when connected. NiMH cells require external charging. Slovenian AAA line conflicts with exact manual and its own four-AA bundle |
| Design | ||
| Display details | Systolic pressure, diastolic pressure and pulse with battery, IHB/AFib and pressure/classification indicators | LCD with ON/OFF control and side 1/3 AFIB/MAM switch; no separate M-button in original S-V9 hardware diagram. IP20 |
| Size | Approximately 96 × 68 × 130 mm (width × height × depth); approximately 240 g excluding batteries; maker lists IP20, without water protection | 152 × 92 × 42 mm; 393 g including batteries, corroborated by exact manual and maker support |
| In the box | UK pack includes UA-611Plus monitor, 22–32 cm cuff, instruction manual and four batteries; product-page box list says AAA but its linked exact manual specifies AA | Slovenian maker lists storage bag, four AA alkaline cells, blood-pressure diary and M-L 22–42 cm cuff. Adapter optional. Product link is Slovenian maker page; support and exact original manual include English |
More comparisons
- A&D Medical UA-611Plus vs Omron M6 Comfort AFib
- A&D Medical UA-611 vs A&D Medical UA-611Plus
- A&D Medical UA-611Plus vs Rossmax AC1000f
- A&D Medical UA-611Plus vs Beurer BM 53
- Beurer BM 53 vs Microlife BP A150 AFIB
- Microlife BP A150 AFIB vs Omron M3 Comfort AFib
- A&D Medical UA-611Plus vs Microlife BP A150-30 AFIB
- A&D Medical UA-611Plus vs Microlife BP A200 AFIB
- A&D Medical UA-611Plus vs Microlife BP A6 PC
- A&D Medical UA-611Plus vs Microlife BP A7 Touch
- A&D Medical UA-611Plus vs Microlife BP B3 AFIB
- A&D Medical UA-611Plus vs Omron M3 Comfort AFib
- A&D Medical UA-611Plus vs Omron M6 Comfort
- A&D Medical UA-611Plus vs Rossmax X5
- A&D Medical UA-611Plus vs Rossmax X9