Beurer BC 21 vs Microlife BP W10 Basic
Better figure, where one is clearly better
| BeurerBC 21 | MicrolifeBP W10 Basic | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Wrist | Wrist |
| Irregular heartbeat detection | Yes | Yes |
| AFib detection or ECG | No | No |
| Method | Automatic oscillometric wrist measurement during deflation. Exact UK item 65046, EAN 4211125650469; SK2 voice-language version. First-sale date unconfirmed | Automatic oscillometric wrist measurement; exact manual describes inflation followed by gradual pressure reduction. No Comfort+ inflation-measurement claim transferred from W3 Comfort. Manufacturer labels it new 2025/26, but exact first-sale date remains unresolved |
| Accuracy | Exact manual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse. Systolic range 60–260 mmHg, diastolic 40–199 mmHg and pulse 40–180 beats/minute; cuff-pressure range 0–300 mmHg is separate | Static pressure accuracy ±3 mmHg and pulse ±5% of readout. Systolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 bpm in current manual; website rounds pulse upper limit to 200. Cuff-pressure display range 0–299 mmHg is separate |
| Clinical validation | Exact manual gives an unspecified clinical-test maximum permissible standard deviation of 8 mmHg, without identifying a published study. Medaval reports no peer-reviewed clinical validation information for BC 21, but incorrectly labels its measurement site upper arm; exact maker manual establishes wrist measurement. Independently established headline validation omitted | Manufacturer product page claims ISO clinical validation and manual claims ESH testing; no exact-model independent validation listing or published trial recovered in bounded searches. Headline validation omitted pending independent evidence |
| Features | Spoken instructions and results in German, English, French, Italian or Turkish for UK item 65046; three volume levels or off. Shared data sheet separately lists item 65040 with English, Spanish, Polish, Arabic and Farsi; those languages are not part of this UK version. Irregular-heartbeat warning, coloured classification scale, date/time, application-error messages and automatic switch-off. No dedicated AFib screening, ECG, movement detection, heart-level position or resting indicator | Ordinary irregular-heartbeat indication, cuff-fit check, arm-movement warning, blood-pressure classification display, low-battery indicator and automatic switch-off after approximately one minute. Website mentions an LED positioning indicator, but exact manual labels a traffic-light display and establishes no separate heart-level positioning sensor; claim remains unresolved. No dedicated AFib algorithm or MAM sequence established |
| Cuff | ||
| Cuff fits | 14–19.5 cm | 13.5–21.5 cm |
| Cuff | Attached wrist cuff for 14–19.5 cm circumference. User must position the cuff at heart level; no electronic position guidance | Pre-formed wrist cuff fits 13.5–21.5 cm; damp cloth and mild soapsuds for cleaning, no machine washing |
| Connections and memory | ||
| Users | 2 | – |
| Readings stored | 60 per user (better) | 30 |
| Bluetooth and app | No | No |
| Memory | Two users with 60 dated readings each; oldest overwritten when full. Average of all stored values and morning (05:00–09:00) or evening (18:00–20:00) readings over the latest seven days. Deletion clears the selected user's entire memory | Stores 30 readings with date/time and displays average of all stored values, excluding readings with suboptimal cuff fit. Oldest reading overwritten when full. Current exact manual explicitly says removing batteries deletes stored measurements; no memory retention borrowed from W1 Basic or W3 Comfort. New reading can be excluded; saved history clears together. No separate user selector established |
| App | No Bluetooth or connection from the monitor. Current Scan & Save support captures displayed values using a smartphone camera into beurer HealthManager Pro; not automatic transfer | – |
| Power | Two 1.5 V LR03 AAA alkaline batteries; approximately 200 measurements depending on blood pressure and inflation pressure. Battery operation only | Two 1.5 V AAA alkaline batteries; approximately 700 measurements with new cells. NiMH rechargeable cells supported but charged externally. Product page and manual agree on two cells; support table incorrectly gives four |
| Design | ||
| Display details | Non-illuminated LCD with black digits, maker size 3.8 × 4.5 cm. Physical membrane memory and start/stop controls; low-battery indicator. Manual IP22 | LCD with ON/OFF, memory and time controls plus separate classification display; IP22 in current exact manual |
| Size | Exact manual and September 2021 data sheet give 87 × 67 × 31 mm (L × W × H) and approximately 105 g without batteries; data sheet specifies cuff included. Current product table instead gives 87 × 67 × 70 mm and 187 g excluding packaging, with weighing conditions unspecified; differences unresolved | Exact operating manual gives 76 × 67 × 29 mm and 127 g including batteries. European product/support tables instead give 131 × 90 × 60.5 mm; larger figures are not adopted as body dimensions or guessed to be packaging |
| In the box | Monitor with attached cuff, two AAA batteries, storage box and instructions. Current product page also lists a blood pressure record | Monitor, storage box, two AAA alkaline batteries and blood-pressure diary listed by manufacturer; operating instructions available |
More comparisons
- Beurer BC 21 vs Microlife BP W1 Basic
- Beurer BC 21 vs Microlife BP W2 Slim
- Beurer BC 27 vs Microlife BP W10 Basic
- Beurer BC 28 vs Microlife BP W10 Basic
- Beurer BC 21 vs Microlife BP W3 Comfort
- Beurer BC 21 vs Microlife BP W70
- Beurer BC 21 vs Microlife BP W90
- Beurer BC 21 vs Omron RS1
- Beurer BC 21 vs Omron RS2
- Beurer BC 21 vs Omron RS4