Beurer BM 81 easyLock vs Microlife BP B1 BT

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BM 81 easyLock compared with BP B1 BT
BeurerBM 81 easyLockMicrolifeBP B1 BT
Overview
Released––
Measuring
TypeUpper armUpper arm
Clinically validated–Yes
Irregular heartbeat detectionYesYes
AFib detection or ECGNoNo
MethodAutomatic oscillometric upper-arm measurement during inflation; exact BM 81 easyLock item 65511. First on-sale date unconfirmedAutomatic oscillometric upper-arm measurement during deflation; technical model BPHJA2-B, reference BP B1 BT. Gentle+ optimises inflation pressure and deflation speed. First on-sale date unresolved
AccuracyMaker display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse; systolic range 50–250 mmHg, diastolic 30–200 mmHg and pulse 40–180 beats/minute. Cuff-pressure range 0–300 mmHg is not the systolic measuring rangeStatic pressure accuracy ±3 mmHg and pulse ±5% of readout; systolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 beats/minute. Cuff-pressure display 0–299 mmHg is separate
Clinical validationMaker claims clinical validation and ESH recognition; Medaval's April 2025 list places this model among monitors with internal or unpublished validation. Independent published exact-device validation not established; headline validation flag omittedExact STRIDE BP listing approves equivalence to BP B3 AFIB, whose adult ISO 81060-2:2018 validation study was published in 2021; cuff 22–42 cm. This is approved equivalence rather than a separate exact-model trial. Manufacturer UK page also cites inherited BIHS A/A claim
FeaturesCuff-fit check, irregular-heartbeat warning, coloured LED risk indicator, date/time and automatic switch-off. No ECG or haemodynamic resting indicator; not a continuous ambulatory monitorOrdinary irregular-heartbeat indication, cuff-fit check, arm-movement and cuff-pressure error warnings, three-level reading scale, low-battery indication and automatic switch-off after approximately one minute. No dedicated AFib algorithm or automatic triple-measurement sequence established by exact instructions
Cuff
Cuff fits24–40 cm22–42 cm
CuffIntegrated easyLock upper-arm cuff with press-and-turn adjustment for 24–40 cm circumference; no separate tube or hook-and-loop fastenerSupplied soft non-washable M–L cuff fits 22–42 cm; compatible cuff and connector required. Replacement cuffs must use the supplied matching connector. Spot-clean cuff with damp cloth; no machine or dishwasher washing
Connections and memory
Users2–
Readings stored120 per user (better)30
Bluetooth and appYesYes
MemoryTwo users with 120 dated readings each; oldest overwritten when full. Averages of all saved readings and morning/evening readings from the latest seven days. All readings saved to a user memory; deletion clears that user's entire memory, without individual-reading deletion or guest modeStores up to 30 readings with date/time and displays their average; oldest overwritten when full. New reading can be excluded, but saved history clears as a whole. Memory erased on battery removal. Manual pairing text refers to user selection 1/2 but exact controls have no user selector or separate per-user capacity; headline user count omitted
AppBluetooth Low Energy transfer to beurer HealthManager Pro; can be used without the app. Current instructions specify compatibility with Bluetooth 4.0 smartphones/tablets, not an exact radio-generation claimBluetooth 5.0 (European manufacturer page), Bluetooth Low Energy (manual), transfers readings to Microlife Connected Health+ on Android/iOS. Wireless transfer starts after measurement or with Bluetooth/time button and switches off after two minutes without a connection. USB-C is a power input, not an established wired data connection
PowerFour 1.5 V AAA LR03 batteries included; approximately 300 measurements depending on blood pressure and inflation pressure. Battery replacement retains readings but requires resetting date/time. ETM's AA battery table conflicts with the exact maker manual and PC-WELT's AAA descriptionFour 1.5 V LR3 AAA alkaline batteries, approximately 400 measurements. Rechargeable NiMH cells supported but must be removed and charged externally. Optional USB-C adapter model DSA-5PF21-05, 100–240 V input and 5 V 1 A 5 W output; batteries are not charged by the monitor
Design
Display detailsNon-illuminated LCD with white digits and coloured LED classification indicator; maker display dimensions 3.4 × 7.9 cm. Battery-level and cuff-fit indications; IP22LCD, ON/OFF, memory and Bluetooth/time controls; IP21 according to current exact manual
SizeCurrent manual gives 143 × 91 × 22 mm (L × W × H), approximately 280 g with cuff but without batteries. Web table reverses length/width and gives 327 g excluding packaging without the same battery condition130 × 93.5 × 52 mm; 240 g including batteries
In the boxMonitor with integrated 24–40 cm cuff, four AAA LR03 batteries, instructions and quick-start guide; no storage bag. UK product page additionally lists a blood pressure recordMonitor, soft 22–42 cm cuff and four AAA batteries listed by UK manufacturer. Mains adapter optional

More comparisons