Photo 1 of 5
Rossmax X9 BT · Photo source
X9 BT
A Bluetooth upper-arm monitor with automatic and stethoscope-assisted measurement, 90-reading memory and three supplied cuff sizes.
- Type
- Upper arm
- Users
- 1
- App
- Yes
- Validated
- –
Product links
Specifications
Report a wrong or missing specMeasuring
- Type
- Upper arm
- Irregular heartbeat detection
- Yes
- AFib detection or ECG
- Yes
- Method
- Automatic oscillometric measurement or auscultatory mode using a separate stethoscope and marker button. Intended for adults at home and in clinical settings. First-sale date unresolved
- Accuracy
- Pressure 30–260 mmHg; pulse 40–199 bpm. Maker accuracy ±3 mmHg pressure and ±5% pulse
- Clinical validation
- Maker carries ESH/BIHS claims. Medaval's exact X9 BT entry says peer-reviewed validation evidence and protocol-compliant equivalence remain unproven; non-Bluetooth X9's registry listing is not transferred
- Features
- PARR flags AFib, premature beats, tachycardia above 100 bpm and bradycardia below 55 bpm, including combined findings. Pulse screening, not ECG. Cuff-fit, movement and hypertension indicators
Cuff
- Cuff fits
- 16–46 cm
- Cuff
- Supplied latex-free small 16–26 cm, medium 24–36 cm and large 34–46 cm cuffs; overall range requires choosing the appropriate cuff. Cuff holder
Connections and memory
- Users
- 1
- Readings stored
- 90
- Bluetooth and app
- Yes
- Memory
- Single 90-reading history with date/time for automatic measurements; oldest replaced when full. Auscultatory readings are not included in this stored automatic-mode history
- App
- Bluetooth low energy sends completed readings to Rossmax Healthstyle after pairing; stored automatic-mode readings can also be transferred. Auscultatory systolic/diastolic readings transfer when marked with the button. Guide documents an optional specially designed USB cable for PC transfer, but Rossmax now says BP Manager software is discontinued and no longer available; monitor remains independently usable
- Power
- Four AA alkaline cells, DC 6 V; supplied 6 V mains adaptor rated at least 600 mA. Automatic shutdown after one minute without key operation
Design
- Display details
- Backlit 99 × 115 mm LCD with mode and pulse-arrhythmia flags; IP21
- Size
- 173 × 115 × 64.5 mm (L × W × H); maker guide lists 410 g without batteries
- In the box
- Batteries, small/medium/large cuffs, 6 V adaptor, instruction manual and zipper bag; PC cable optional. Stethoscope not listed as supplied
Pros and cons
Pros
- Automatic and auscultatory modes
- Three cuff sizes supplied
- Bluetooth reading transfer
Cons
- Stethoscope required for auscultatory measurement
- Single shared reading history
- PC software discontinued
Closest rivals
Model | Type | Users | App | Validated | Released |
|---|---|---|---|---|---|
| Upper arm | 2 | Yes | – | – | |
| Upper arm | 2 | Yes | – | – | |
| Upper arm | 1 | Yes | Yes | – |
Sources
- rossmax.com/en/products/monitoring/blood-pressure-monitors.html
- rossmax.com/downloads/IBs/OBM_IB_X9%282%29NA18_EN.pdf
- rossmax.com/downloads/Epaper/OBM_Epaper_X9%282%29NA18_EN.pdf
- rossmax.com/en/rossmax-downloads.html
- medaval.ie/resources/EN/devices/Rossmax-X9-BT.html
Photo: rossmax.com
- X9 BT manufacturer low-angle studio photograph of the monitor and cuff: rossmax.com/images/products/X9BT/X9BT_PP_006.png
- X9 BT manufacturer studio kit photograph showing the three supplied cuffs: rossmax.com/images/products/X9BT/X9BT_PP_005.png
- Maker’s shared X9-family data and power sockets, linked from the X9 BT page: rossmax.com/images/products/X9/X9_PP_005.jpg
- X9 BT, monitor without cuff: rossmax.com/images/products/X9BT/X9BT_PP_001_5.png