Beurer BC 27 vs Microlife BP W1 Basic
Better figure, where one is clearly better
| BeurerBC 27 | MicrolifeBP W1 Basic | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Wrist | Wrist |
| Irregular heartbeat detection | Yes | Yes |
| AFib detection or ECG | No | No |
| Method | Automatic oscillometric wrist measurement during deflation. Exact UK item 65904, EAN 4211125659042. Current page heading and download labels identify BC 27 despite BC 28 in its SEO title and range tile; linked manual identifies type BC 27, technical model BC 28. First-sale date unconfirmed | Automatic oscillometric wrist measurement during inflation, explicitly described in current exact manual despite Gentle+ marketing name. First on-sale date unresolved |
| Accuracy | Manual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse. Systolic range 60–260 mmHg, diastolic 40–199 mmHg and pulse 40–180 beats/minute; cuff-pressure range 0–300 mmHg is separate | Static pressure accuracy ±3 mmHg, pulse ±5% of readout; systolic 60–255 mmHg and diastolic 40–200 mmHg. Current manual pulse range 40–199 takes precedence over website 40–200; cuff-pressure display range 0–299 mmHg is separate |
| Clinical validation | Exact manual states a maximum permissible clinical-test standard deviation of 8 mmHg, but does not identify a published validation study. Medaval reports no peer-reviewed clinical validation information for BC 27; consumer laboratory tests do not establish clinical validation. Headline validation omitted | Manufacturer website claims BIHS A/A and manual states ESH testing; Medaval exact-model record reports no peer-reviewed clinical validation evidence. Headline validation omitted rather than treating manufacturer claim as an independently established trial |
| Features | Irregular-heartbeat warning, coloured classification scale, date/time, application-error messages and automatic switch-off after approximately one minute. No dedicated AFib screening, ECG, movement detection, heart-level position indicator, resting indicator or voice output | LED positioning indicator helps set wrist at heart level; inflation starts after 5–8 seconds. Ordinary irregular-heartbeat indication, traffic-light scale, cuff-fit and arm-movement indicators, low-battery warning and automatic switch-off after approximately one minute. No dedicated AFib algorithm or MAM sequence established |
| Cuff | ||
| Cuff fits | 14–19.5 cm | 13.5–21.5 cm |
| Cuff | Fixed wrist cuff for 14–19.5 cm circumference; exact manual requires the supplied attached cuff. Heart-level positioning is a user instruction, not an electronic position sensor | Attached wrist cuff fits 13.5–21.5 cm according to current manual; clean marks with damp cloth and mild detergent |
| Connections and memory | ||
| Users | 2 | – |
| Readings stored | 60 per user | 60 |
| Bluetooth and app | No | No |
| Memory | Two users with 60 dated readings each; oldest overwritten when full. Average of all stored readings and seven-day morning (05:00–09:00) or evening (18:00–20:00) averages. Selected user memory can be cleared | Stores 60 readings with date/time, oldest overwritten when full; readings retained on battery replacement while clock resets. New result can be excluded; saved history clears as a whole rather than individually. Manual says per user but exact labelled controls/display establish no separate user selection, so no headline user count |
| App | No Bluetooth or connection from the monitor. Scan & Save captures the displayed values using a phone camera into beurer HealthManager Pro; not automatic wireless transfer | – |
| Power | Two 1.5 V LR03 AAA batteries. Exact linked manual claims approximately 170 measurements depending on blood pressure and inflation pressure; current product table says approximately 300. No mains operation established | Two 1.5 V AAA alkaline batteries, approximately 320 measurements with new cells. Current support table incorrectly lists four AAA cells; product page and exact manual both specify two. NiMH rechargeable cells supported but charged externally, not inside the monitor |
| Design | ||
| Display details | LCD with black digits; current maker table gives 3.5 × 4.0 cm and claims illumination, while FAZ Kaufkompass's exact tested BC 27 has an unlit display. Illumination remains unresolved rather than assuming a hardware revision. Physical start/stop and memory controls, low-battery indicator; manual IP22 | LCD with ON/OFF, memory and time controls; LED positioning indicator. IP22 in current exact manual |
| Size | Exact linked manual 84 × 60 × 29 mm (L × W × H), approximately 92 g with cuff without batteries. Current web/data sheet instead gives 60 × 84 × 55 mm and approximately 171 g excluding packaging, without equivalent weighing conditions; conflict unresolved | 73 × 66 × 38 mm; 124 g including batteries |
| In the box | Monitor with attached cuff, two AAA batteries, storage box and instructions. Current product page and data sheet also list a blood pressure record | Monitor, storage box, two AAA alkaline batteries and blood-pressure diary listed by European manufacturer; official instructions available |
More comparisons
- Beurer BC 21 vs Microlife BP W1 Basic
- Beurer BC 27 vs Microlife BP W10 Basic
- Beurer BC 27 vs Microlife BP W2 Slim
- Beurer BC 28 vs Microlife BP W1 Basic
- Beurer BC 27 vs Microlife BP W3 Comfort
- Beurer BC 27 vs Microlife BP W70
- Beurer BC 27 vs Microlife BP W90
- Beurer BC 27 vs Omron RS1
- Beurer BC 27 vs Omron RS2
- Beurer BC 27 vs Omron RS4