Beurer BC 27 vs Microlife BP W3 Comfort

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BC 27 compared with BP W3 Comfort
BeurerBC 27MicrolifeBP W3 Comfort
Overview
Released––
Measuring
TypeWristWrist
Clinically validated–Yes
Irregular heartbeat detectionYesYes
AFib detection or ECGNoNo
MethodAutomatic oscillometric wrist measurement during deflation. Exact UK item 65904, EAN 4211125659042. Current page heading and download labels identify BC 27 despite BC 28 in its SEO title and range tile; linked manual identifies type BC 27, technical model BC 28. First-sale date unconfirmedAutomatic oscillometric wrist measurement during inflation with Comfort+; exact current European manual identifies BP3KB1-3E. First on-sale date unresolved
AccuracyManual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse. Systolic range 60–260 mmHg, diastolic 40–199 mmHg and pulse 40–180 beats/minute; cuff-pressure range 0–300 mmHg is separateStatic pressure accuracy ±3 mmHg and pulse ±5% of readout. Current European manual gives systolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 bpm; older UK manual gives 20–280 mmHg blood pressure and 40–200 bpm. Cuff-pressure display 0–299 mmHg is separate
Clinical validationExact manual states a maximum permissible clinical-test standard deviation of 8 mmHg, but does not identify a published validation study. Medaval reports no peer-reviewed clinical validation information for BC 27; consumer laboratory tests do not establish clinical validation. Headline validation omittedSTRIDE BP approves equivalence to W2 Slim, whose 2016 adult validation used ESH-IP 2010 and ANSI/AAMI/ISO 2013; not a separate published W3 Comfort trial. STRIDE lists a 13.5–23 cm cuff. Manufacturer BIHS A/A marketing and current manual ESH/ISO wording are retained as claims, not additional independent trials
FeaturesIrregular-heartbeat warning, coloured classification scale, date/time, application-error messages and automatic switch-off after approximately one minute. No dedicated AFib screening, ECG, movement detection, heart-level position indicator, resting indicator or voice outputOrdinary irregular-heartbeat indication, cuff-fit check, arm-movement warning, single-colour blood-pressure classification scale, low-battery indicator and automatic switch-off after approximately one minute. No dedicated AFib algorithm, MAM sequence or electronic wrist-position sensor established
Cuff
Cuff fits14–19.5 cm–
CuffFixed wrist cuff for 14–19.5 cm circumference; exact manual requires the supplied attached cuff. Heart-level positioning is a user instruction, not an electronic position sensorUK product page and its EN-FR 4119 manual give 13.5–23 cm and a soft D-ring wrist cuff. European August 2026 manual instead gives 13.5–21.5 cm despite the European product page still advertising 23 cm. Headline cuff range omitted pending exact supplied revision confirmation; no inferred second hardware model
Connections and memory
Users2–
Readings stored60 per user (better)30
Bluetooth and appNoNo
MemoryTwo users with 60 dated readings each; oldest overwritten when full. Average of all stored readings and seven-day morning (05:00–09:00) or evening (18:00–20:00) averages. Selected user memory can be clearedStores 30 readings with date/time and displays average of all stored readings, excluding suboptimal cuff fit; not specifically the last three. Oldest overwritten when full. Readings retained on cell replacement, clock reset. New reading can be excluded; saved history clears together. Exact controls establish no separate user selector
AppNo Bluetooth or connection from the monitor. Scan & Save captures the displayed values using a phone camera into beurer HealthManager Pro; not automatic wireless transfer–
PowerTwo 1.5 V LR03 AAA batteries. Exact linked manual claims approximately 170 measurements depending on blood pressure and inflation pressure; current product table says approximately 300. No mains operation establishedTwo 1.5 V AAA alkaline cells; approximately 320 measurements with new cells. NiMH cells supported but must be removed and charged externally
Design
Display detailsLCD with black digits; current maker table gives 3.5 × 4.0 cm and claims illumination, while FAZ Kaufkompass's exact tested BC 27 has an unlit display. Illumination remains unresolved rather than assuming a hardware revision. Physical start/stop and memory controls, low-battery indicator; manual IP22LCD with ON/OFF, memory and time controls. Older UK manual gives IP20; current European August 2026 revision gives IP22. Revision conflict retained rather than assuming an upgrade applies to every unit
SizeExact linked manual 84 × 60 × 29 mm (L × W × H), approximately 92 g with cuff without batteries. Current web/data sheet instead gives 60 × 84 × 55 mm and approximately 171 g excluding packaging, without equivalent weighing conditions; conflict unresolved81 × 65 × 21 mm; 112 g including batteries
In the boxMonitor with attached cuff, two AAA batteries, storage box and instructions. Current product page and data sheet also list a blood pressure recordMonitor, storage case, two AAA alkaline batteries, instructions and blood-pressure diary. Current range listing establishes catalogue presence, not worldwide stock or first-sale date