Beurer BC 28 vs Microlife BP W70

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BC 28 compared with BP W70
BeurerBC 28MicrolifeBP W70
Overview
Released––
Measuring
TypeWristWrist
Irregular heartbeat detectionYesYes
AFib detection or ECGNoNo
MethodAutomatic oscillometric wrist measurement during deflation. Exact white UK item 65044, EAN 4211125650445; BC 27 dark-grey item 65904 has a separate maker product record. First-sale date unconfirmedOscillometric wrist measurement during inflation, marketed as Comfort+ and Quiet Inflation. This is the non-Bluetooth BP W70, separately listed from W70 BT by the Asia-Pacific manufacturer; no inferred successor link. First on-sale date unresolved
AccuracyExact manual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse. Systolic range 60–260 mmHg, diastolic 40–199 mmHg and pulse 40–180 beats/minute; cuff-pressure range 0–300 mmHg is separateSystolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 bpm in maker product/support. Static pressure accuracy ±3 mmHg and pulse ±5% in exact ROA 0922 manual. Cuff-pressure display 0–299 mmHg is separate
Clinical validationManual gives an unspecified clinical-test maximum permissible standard deviation of 8 mmHg for systolic and diastolic pressure, without identifying a published study. Medaval reports no peer-reviewed clinical validation information for BC 28; headline independently established validation omittedExact manual states testing according to ESH and ISO 81060-2:2013. No exact published clinical study or independent validation-registry assessment recovered in this pass; headline validation remains unresolved rather than transferring W70 BT evidence
FeaturesIrregular-heartbeat warning, coloured classification scale, date/time, application-error messages and automatic switch-off after approximately one minute. No dedicated AFib screening, ECG, movement detection, heart-level position indicator, resting indicator or voice outputSmartMAM takes two consecutive measurements with a 15-second interval and displays an analysed result; a questionable reading can trigger another measurement. This differs from three-reading claims for some W70 BT versions. Ordinary pulse-arrhythmia indication, cuff-fit and movement checks, traffic-light classification and automatic switch-off after approximately one minute; no dedicated AFib screening
Cuff
Cuff fits14–19.5 cm13.5–21.5 cm
CuffAttached wrist cuff for 14–19.5 cm circumference. User must position the cuff at heart level; no electronic position guidanceWrist circumference 13.5–21.5 cm. Approximate five-level wrist-position guidance and blue LED; enabled guidance can begin measurement after the correct position is retained for approximately two seconds. Manual qualifies limitations with individual physique and allows the guidance to be disabled
Connections and memory
Users22
Readings stored60 per user99 per user (better)
Bluetooth and appNoNo
MemoryTwo users with 60 dated readings each; oldest overwritten when full. Average of all stored values and morning (05:00–09:00) or evening (18:00–20:00) readings over the latest seven days. Deletion clears the selected user's entire memoryTwo users selected with the time button, 99 dated readings each. Recall first shows the average of stored values, excluding suboptimal cuff-fit readings, then individual results. Oldest overwritten when full. Stored readings survive cell replacement while settings and clock need resetting. Newly displayed reading can be excluded; previously stored individual readings cannot be deleted separately
AppNo Bluetooth or connection from the monitor. Current Scan & Save support captures the displayed values using a smartphone camera into beurer HealthManager Pro; not automatic transfer–
PowerTwo 1.5 V LR03 AAA alkaline batteries; exact manuals claim approximately 170 measurements depending on blood pressure and inflation pressure, while the current product table says approximately 300. Battery operation onlyTwo 1.5 V AAA alkaline cells, approximately 300 measurements with new batteries in exact ROA 0922 manual. NiMH rechargeable cells supported with external charging; no internal charging or adapter interface established
Design
Display detailsNon-illuminated LCD with black digits, maker size 3.5 × 4.0 cm. Physical start/stop and memory controls; low-battery indicator. Manual IP22LCD with ON/OFF, memory and time buttons; IP22 in exact manual. Device service life stated as five years or 10,000 measurements, accessories two years; not battery endurance
SizeExact English and March 2022 multilingual manuals and January 2019 data sheet give 84 × 60 × 29 mm (L × W × H), approximately 92 g with cuff without batteries. Current product table instead gives 60 × 84 × 55 mm and 171 g excluding packaging, with weighing conditions unspecified; differences unresolved82 × 64 × 13.9 mm; 78 ±5 g including batteries in exact maker product, support and manual
In the boxMonitor with attached cuff, two AAA batteries, storage box and instructions. Current product page also lists a blood pressure record. Dedicated English manual's footer is September 2019 despite its download filename using March 2022; newer multilingual manual has the same core specificationsExact included accessory inventory not established by sources read. Resources are the English Asia-Pacific maker product/support and exact English manual; no unsupported UK availability claim