Beurer BM 23 vs Microlife BP A2 Classic

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BM 23 compared with BP A2 Classic
BeurerBM 23MicrolifeBP A2 Classic
Overview
Released––
Measuring
TypeUpper armUpper arm
Clinically validated–Yes
Irregular heartbeat detectionYesYes
AFib detection or ECGNoNo
MethodAutomatic oscillometric upper-arm measurement during deflation. Exact UK item 60493, EAN 4211125604936; manual model BM 23, technical type BM 25. Range tile and SEO title use BM 25, but active product heading, photos, datasheet and instructions identify BM 23. First-sale date unresolvedAutomatic oscillometric upper-arm measurement during deflation, marketed as Gentle+ pressure control. January 2024 exact manual describes pumping followed by gradual pressure reduction; no automatic multi-measurement mode. First actual on-sale date unresolved
AccuracyManual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse; systolic range 50–280 mmHg, diastolic 30–200 mmHg and pulse 40–199 beats/minute. Maximum cuff pressure 300 mmHg is separateJanuary 2024 E-V10 manual gives systolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 bpm; static pressure ±3 mmHg and pulse ±5%. Hungarian maker page instead gives blood pressure 20–280 mmHg and pulse 40–200 bpm; global support gives separate systolic/diastolic ranges but pulse 40–200. Cuff-pressure display 0–299 is separate
Clinical validationCurrent manual claims clinical testing to ISO 81060-2. Medaval lists a Stride BP derivative recommendation based on equivalence to BM 28 via Globalcare GCE603, but says unpublished equivalence results cannot be verified to MDR requirements. Independently established exact-device validation remains unresolved; headline flag omittedSTRIDE BP lists A2 Classic as equivalent to BP A100, for adults and diabetes with 22–42 and 32–52 cm cuffs. Headline reflects registry acceptance by equivalence, not a separate exact-model trial. Exact manual cites BP 3BT0-A technology; broad manufacturer special-population claims are not assigned as independently established exact-model trials
FeaturesIrregular-heartbeat warning, coloured classification scale, electronic cuff-position check, automatic switch-off and battery indicator. No ECG, dedicated AFib screening, resting indicator or voice outputOrdinary irregular-heartbeat indication, threshold classification scale with 140/90 mark, low-battery indicator and automatic switch-off after approximately one minute. No dedicated AFib screening or app
Cuff
Cuff fits22–42 cm22–42 cm
CuffSupplied universal cuff for 22–42 cm upper-arm circumference, replacement item 110.031; conventional hook-and-loop cuff rather than easyClip or easyLockSupplied conical M-L cuff fits 22–42 cm in European and Thai maker pages. Manual separately lists optional S 17–22, M 22–32, L 32–42 and L-XL 32–52 cm; these are not all supplied
Connections and memory
Readings stored–30
Bluetooth and appNoNo
MemoryNo stored reading history, user profiles, date/time or calculated averages established by the product table or complete device operation section. Manual's isolated statement about resetting date/time and retaining saved measurements conflicts with those sections; not treated as an implemented featureLast 30 readings stored, with average displayed first on memory recall followed by individual values. Oldest overwritten when full; newly displayed reading can be excluded and stored history clears together. No clock-setting control, dated-reading recall or user-selector operation established. Hungarian page's translated 30-day wording is not used as a storage-duration claim
AppNo Bluetooth connection from the monitor. Scan & Save captures the display with a smartphone camera into beurer HealthManager Pro rather than automatically transmitting measurements–
PowerFour 1.5 V LR03 AAA batteries included; manual approximately 300 measurements depending on blood pressure and inflation pressure. USB-C input supports optional 5 V DC / 1 A mains operation, not battery charging or a demonstrated data connection. Manual lists EU adapter 072.78 and UK adapter 072.79; remove batteries for mains operationFour 1.5 V AA alkaline cells, approximately 920 measurements with new batteries in January 2024 manual. Microlife DC 6 V, 600 mA adapter supported through dedicated socket; no battery current consumed when connected. NiMH cells supported with external charging, not charging inside the monitor
Design
Display detailsNon-illuminated LCD with black digits, maker 4.4 × 5.3 cm; physical START/STOP button. Manual IP21LCD with ON/OFF and memory buttons, cuff and mains-adapter sockets; no time button. IP20 in January 2024 manual
Size120 × 90 × 40 mm (L × W × H). Manual and datasheet approximately 318 g with cuff and without batteries; current product table's 362 g excluding packaging has unspecified weighing conditions120 × 87 × 59 mm; 375 g including batteries in exact global support, European maker page and manual
In the boxMonitor, universal upper-arm cuff, four AAA batteries and instructions according to the datasheet and manual. Product page additionally lists a blood pressure record; mains adapter sold separatelyEuropean maker lists four AA alkaline cells, blood-pressure diary and M-L 22–42 cm cuff, with adapter optional. Thai regional bundle also includes textile storage bag, instructions and DC6V600mA adapter; this does not establish adapter inclusion in every region. Product link is Hungarian maker page; support and linked manual include English