Beurer BM 23 vs Microlife BP A3 PC

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BM 23 compared with BP A3 PC
BeurerBM 23MicrolifeBP A3 PC
Overview
Released––
Measuring
TypeUpper armUpper arm
Clinically validated–Yes
Irregular heartbeat detectionYesYes
AFib detection or ECGNoNo
MethodAutomatic oscillometric upper-arm measurement during deflation. Exact UK item 60493, EAN 4211125604936; manual model BM 23, technical type BM 25. Range tile and SEO title use BM 25, but active product heading, photos, datasheet and instructions identify BM 23. First-sale date unresolvedOscillometric upper-arm measurement during deflation. Side switch selects a single measurement or MAM mode, which analyses three readings with 15-second pauses into one displayed result. Individual MAM results are not displayed; a fourth measurement is taken if one is questionable. First actual on-sale date unresolved
AccuracyManual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse; systolic range 50–280 mmHg, diastolic 30–200 mmHg and pulse 40–199 beats/minute. Maximum cuff pressure 300 mmHg is separateExact April 2020 English/Greek guide gives blood pressure 20–280 mmHg, pulse 40–200 bpm, static accuracy ±3 mmHg and pulse accuracy ±5%. Current global support instead separates systolic 60–255 and diastolic 40–200 mmHg; range difference remains explicit. Cuff-pressure display 0–299 mmHg is separate
Clinical validationCurrent manual claims clinical testing to ISO 81060-2. Medaval lists a Stride BP derivative recommendation based on equivalence to BM 28 via Globalcare GCE603, but says unpublished equivalence results cannot be verified to MDR requirements. Independently established exact-device validation remains unresolved; headline flag omittedSTRIDE BP lists an exact-model diabetes study using ANSI/AAMI/ISO 2013, published in 2018, and adult acceptance by equivalence to BP A100. Published Beime study abstract reports 85 diabetic participants aged 56–88 and passing both protocol criteria; this does not establish every population in broad maker claims. Registry lists 22–42 and 32–52 cm validated cuffs
FeaturesIrregular-heartbeat warning, coloured classification scale, electronic cuff-position check, automatic switch-off and battery indicator. No ECG, dedicated AFib screening, resting indicator or voice outputPulse Arrhythmia Detection indicates pulse irregularities, not a separate AFib algorithm or ECG. Cuff-fit and arm-movement indicators, coloured classification scale and low-battery warning; automatic switch-off after approximately one minute
Cuff
Cuff fits22–42 cm–
CuffSupplied universal cuff for 22–42 cm upper-arm circumference, replacement item 110.031; conventional hook-and-loop cuff rather than easyClip or easyLockExact guide lists S 17–22, M 22–32, M-L 22–42, L 32–42 and L-XL 32–52 cm cuffs and optional preformed Easy cuffs. Recovered support record does not establish the supplied size, so headline cuff range is omitted. Guide describes removable bladder and hand-washable cuff cover; do not wash bladder
Connections and memory
Users–2
Readings stored–99 per user
Bluetooth and appNoNo
MemoryNo stored reading history, user profiles, date/time or calculated averages established by the product table or complete device operation section. Manual's isolated statement about resetting date/time and retaining saved measurements conflicts with those sections; not treated as an implemented featureUser switch selects one of two separate histories, each storing 99 dated readings. First recall shows average of all stored values, followed by individual history. Oldest overwritten when full. All values for selected user can be cleared, but individual stored values cannot; a newly displayed result can be excluded before storage. History survives battery replacement while clock requires resetting
AppNo Bluetooth connection from the monitor. Scan & Save captures the display with a smartphone camera into beurer HealthManager Pro rather than automatically transmitting measurementsUSB PC link to Microlife Blood Pressure Analyzer+ using Mini-B 5-pin connector. Computer controls device during connection. No Bluetooth or mobile-app transfer. Current software compatibility must be checked separately rather than inferred from old CDs
PowerFour 1.5 V LR03 AAA batteries included; manual approximately 300 measurements depending on blood pressure and inflation pressure. USB-C input supports optional 5 V DC / 1 A mains operation, not battery charging or a demonstrated data connection. Manual lists EU adapter 072.78 and UK adapter 072.79; remove batteries for mains operationFour 1.5 V AA alkaline cells; approximately 920 measurements with new batteries, without a separate MAM-cycle count. Optional Microlife DC 6 V, 600 mA adapter stops battery-current use when connected. NiMH cells supported but require external charging
Design
Display detailsNon-illuminated LCD with black digits, maker 4.4 × 5.3 cm; physical START/STOP button. Manual IP21LCD with date/time and user indicator, ON/OFF, memory and time buttons, separate user and 1/3 MAM switches. USB and DC adapter sockets. IP20
Size120 × 90 × 40 mm (L × W × H). Manual and datasheet approximately 318 g with cuff and without batteries; current product table's 362 g excluding packaging has unspecified weighing conditions143 × 85 × 58 mm; 368 g including batteries in exact guide and global support
In the boxMonitor, universal upper-arm cuff, four AAA batteries and instructions according to the datasheet and manual. Product page additionally lists a blood pressure record; mains adapter sold separatelyExact guide allows regional packages without cable or download voucher; complete original supplied bundle unresolved. Verified global archived specification/support record serves both resource types. English guide includes Greek regional guarantee wording; this is not a worldwide warranty claim