Beurer BM 38 vs Microlife BP 3AS1-2
Better figure, where one is clearly better
| BeurerBM 38 | MicrolifeBP 3AS1-2 | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Upper arm | Upper arm |
| Clinically validated | – | Yes |
| Irregular heartbeat detection | Yes | No |
| AFib detection or ECG | No | No |
| Method | Automatic oscillometric upper-arm measurement during inflation. UK item 10361, EAN 4211125103613; manual model BM 38, technical type BM 48/1. Announced at IFA on 6 September 2024 with planned German availability from October 2024; first actual sale unresolved | Semi-automatic oscillometric upper-arm measurement with capacitive pressure sensor. Hand inflation using pump integrated into handle; automatic deflation and digital pressure/pulse results. O/I button interrupts measurement and releases cuff pressure. Actual first retail date unresolved |
| Accuracy | Manual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse; systolic range 50–280 mmHg, diastolic 30–200 mmHg and pulse 40–199 beats/minute. Maximum cuff pressure 300 mmHg is separate | February 2024 exact guide gives systolic/diastolic range 20–280 mmHg, pulse 40–200 bpm, 1 mmHg pressure resolution, static pressure accuracy ±3 mmHg and pulse accuracy ±5% of reading |
| Clinical validation | Current manual claims clinical testing to ISO 81060-2. Independent exact-device validation not established in the source pass; headline flag omitted | Nathan and colleagues, Blood Pressure Monitoring 2015, tested exact Microlife 3AS1-2 in 45 pregnant women including 15 with pre-eclampsia. Overall B/A grade under the British Hypertension Society protocol and passed the study's ISO mean-difference standard; pregnancy-only subgroup A/A. These are study-specific findings, not evidence for every population named in generic maker text |
| Features | Irregular-heartbeat warning, coloured classification scale, cuff-position check and HSD resting indicator. Automatic switch-off after approximately one minute, battery indicator and START/STOP button. No ECG, dedicated AFib screening or voice output | Manual inflation, automatic measurement/deflation, low-battery and inflation indicators. Flashing heart indicates detected pulse, not irregular-heartbeat screening. No AFib detection, ECG or automatic triple-measurement mode |
| Cuff | ||
| Cuff fits | 22–42 cm | 22–32 cm |
| Cuff | Supplied conventional universal cuff for 22–42 cm upper-arm circumference, replacement item 110.031; no easyClip or easyLock mechanism | Supplied M 22–32 cm cuff; optional L 32–42 cm. Single-tube cuff with T-connector attaches at the back of the device |
| Connections and memory | ||
| Users | – | 1 |
| Readings stored | – | 1 |
| Bluetooth and app | No | No |
| Memory | Product table says no calculated averages. No reading-recall, user-selection or clock-setting controls established by the complete operating instructions. Website date/time claim and manual's isolated saved-measurement/user paragraphs conflict with the controls; stored memory and user counts left unresolved | One latest measurement retained until the next measurement or battery removal. Hold O/I for at least three seconds to recall. No dated history, user selector or averaging |
| App | Scan & Save photographs the display into beurer HealthManager Pro. No Bluetooth connection from the monitor; smartphone-camera capture is separate from wireless transmission | No Bluetooth, USB or phone-app transfer |
| Power | Four included 1.5 V LR03 AAA batteries; approximately 300 measurements depending on blood pressure and inflation pressure. USB-C input supports optional 5 V DC / 1 A mains operation, not battery charging. Remove batteries for mains operation; manual lists EU adapter 072.78 and UK adapter 072.79 | Two 1.5 V AAA alkaline cells; approximately 780 measurements using new batteries. Current guide does not recommend 1.2 V rechargeable cells. Generic guarantee wording about an optional adaptor does not establish a mains socket or mains operation |
| Design | ||
| Display details | Non-illuminated LCD with black digits; maker display size 6.2 × 6.2 cm. Manual IP20 | LCD showing systolic, diastolic and pulse readings; single O/I button. IP20 |
| Size | August 2026 manual 128 × 91 × 46 mm (L × W × H), approximately 357 g with cuff and without batteries. Current website and datasheet give 130 × 91 × 46 mm and approximately 385 g excluding packaging, without defining battery/cuff weighing conditions | 68 × 186 × 48.5 mm (width × length × height); 310 g including batteries, excluding cuff in global support table |
| In the box | Monitor, universal 22–42 cm cuff, four AAA batteries and instructions. Mains adapter sold separately | Exact guide establishes monitor with integrated pump and M 22–32 cm cuff; complete regional accessory list unresolved |
More comparisons
- Beurer BM 38 vs Beurer BM 85
- Beurer BM 23 vs Microlife BP 3AS1-2
- Beurer BM 25 vs Microlife BP 3AS1-2
- Beurer BM 27 vs Microlife BP 3AS1-2
- Beurer BM 28 vs Microlife BP 3AS1-2
- Beurer BM 35 vs Microlife BP 3AS1-2
- A&D Medical UA-611 vs Beurer BM 38
- Beurer BM 38 vs Microlife BP 3AG1
- Beurer BM 51 easyClip vs Microlife BP 3AS1-2
- Beurer BM 58 vs Microlife BP 3AS1-2
- A&D Medical UA-611 vs Microlife BP 3AS1-2