A&D Medical UA-611 vs Beurer BM 38
Better figure, where one is clearly better
| A&D MedicalUA-611 | BeurerBM 38 | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Upper arm | Upper arm |
| Irregular heartbeat detection | Yes | Yes |
| AFib detection or ECG | – | No |
| Method | Automatic oscillometric measurement with fuzzy-logic-controlled inflation; designed for adult home use; first on-sale date not established | Automatic oscillometric upper-arm measurement during inflation. UK item 10361, EAN 4211125103613; manual model BM 38, technical type BM 48/1. Announced at IFA on 6 September 2024 with planned German availability from October 2024; first actual sale unresolved |
| Accuracy | Maker specification ±3 mmHg pressure and ±5% pulse; pressure range 0–299 mmHg, systolic 60–279 mmHg, diastolic 40–200 mmHg and pulse 40–180 beats/min | Manual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse; systolic range 50–280 mmHg, diastolic 30–200 mmHg and pulse 40–199 beats/minute. Maximum cuff pressure 300 mmHg is separate |
| Clinical validation | A&D UK claims ESH 2010 validation; Medaval identifies the UK BIHS recommendation as manufacturer-declared equivalence to the UA-651 and says the equivalence study is unpublished. The US Validate BP listing covers UA-611 under ANSI/AAMI/ISO 81060-2:2018 with different cuff options; this does not establish direct validation of the supplied UK cuff configuration | Current manual claims clinical testing to ISO 81060-2. Independent exact-device validation not established in the source pass; headline flag omitted |
| Features | Irregular heartbeat indicator, pressure progress bar, WHO classification indicator, stored-reading average and automatic power-off; irregular heartbeat indication is not a specific AFib diagnosis | Irregular-heartbeat warning, coloured classification scale, cuff-position check and HSD resting indicator. Automatic switch-off after approximately one minute, battery indicator and START/STOP button. No ECG, dedicated AFib screening or voice output |
| Cuff | ||
| Cuff fits | 22–32 cm | 22–42 cm |
| Cuff | Supplied latex-free and metal-free SlimFit cuff fits 22–32 cm; optional large cuff fits 31–45 cm, sold separately; UK manual says this model is not designed for a small cuff | Supplied conventional universal cuff for 22–42 cm upper-arm circumference, replacement item 110.031; no easyClip or easyLock mechanism |
| Connections and memory | ||
| Readings stored | 30 | – |
| Bluetooth and app | No | No |
| Memory | Last 30 blood pressure and pulse measurements, with stored-reading average; no separate user memories listed | Product table says no calculated averages. No reading-recall, user-selection or clock-setting controls established by the complete operating instructions. Website date/time claim and manual's isolated saved-measurement/user paragraphs conflict with the controls; stored memory and user counts left unresolved |
| App | No Bluetooth; readings can be recorded manually | Scan & Save photographs the display into beurer HealthManager Pro. No Bluetooth connection from the monitor; smartphone-camera capture is separate from wireless transmission |
| Power | Four AA 1.5 V batteries; maker claims approximately 700 measurements with LR6 alkaline batteries or 200 with R6P manganese batteries at 180 mmHg and 23°C. UA-611 has no DC socket; the optional adapter in the shared manual applies only to UA-651 | Four included 1.5 V LR03 AAA batteries; approximately 300 measurements depending on blood pressure and inflation pressure. USB-C input supports optional 5 V DC / 1 A mains operation, not battery charging. Remove batteries for mains operation; manual lists EU adapter 072.78 and UK adapter 072.79 |
| Design | ||
| Display details | Systolic pressure, diastolic pressure and pulse shown with battery, irregular heartbeat and pressure/classification indicators | Non-illuminated LCD with black digits; maker display size 6.2 × 6.2 cm. Manual IP20 |
| Size | Approximately 96 × 68 × 130 mm (width × height × depth); approximately 230 g excluding batteries | August 2026 manual 128 × 91 × 46 mm (L × W × H), approximately 357 g with cuff and without batteries. Current website and datasheet give 130 × 91 × 46 mm and approximately 385 g excluding packaging, without defining battery/cuff weighing conditions |
| In the box | UA-611 monitor, 22–32 cm SlimFit cuff, instruction manual and four AA batteries in the UK pack | Monitor, universal 22–42 cm cuff, four AAA batteries and instructions. Mains adapter sold separately |
More comparisons
- Beurer BM 38 vs Beurer BM 85
- A&D Medical UA-611 vs A&D Medical UA-611Plus
- A&D Medical UA-611 vs Beurer BM 54
- A&D Medical UA-611 vs Rossmax AU941f
- A&D Medical UA-611 vs Rossmax AX356f
- A&D Medical UA-611 vs Beurer BM 23
- A&D Medical UA-611 vs Beurer BM 25
- A&D Medical UA-611 vs Beurer BM 27
- A&D Medical UA-611 vs Beurer BM 28
- A&D Medical UA-611 vs Beurer BM 35
- Beurer BM 38 vs Microlife BP 3AG1
- Beurer BM 38 vs Microlife BP 3AS1-2
- A&D Medical UA-611 vs Beurer BM 51 easyClip
- A&D Medical UA-611 vs Beurer BM 58
- A&D Medical UA-611 vs Microlife BP 3AG1
- A&D Medical UA-611 vs Microlife BP 3AS1-2