Beurer BM 54 vs Omron M2+ Connect (HEM-7188T1-LE)
Better figure, where one is clearly better
| BeurerBM 54 | OmronM2+ Connect (HEM-7188T1-LE) | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Upper arm | Upper arm |
| Clinically validated | – | Yes |
| Irregular heartbeat detection | Yes | Yes |
| AFib detection or ECG | – | No |
| Method | Automatic oscillometric measurement on the upper arm during deflation; first on-sale date not established | Automatic oscillometric upper-arm measurement; exact HEM-7188T1-LE. First retail sale date unresolved; manual publication dates are not substituted |
| Accuracy | Maker manual specifies display accuracy of ±3 mmHg for systolic and diastolic pressure and ±5% of the displayed pulse value; this is distinct from an independent clinical validation result | Maker pressure accuracy ±3 mmHg and pulse accuracy ±5% of displayed reading; systolic range 60–260 mmHg, diastolic 40–215 mmHg and pulse 40–180 beats/minute |
| Clinical validation | Beurer states clinical testing to ISO 81060-2 in the current manual and an ESH validation claim in its datasheet | STRIDE BP lists this exact model for adult home use by approved equivalence to Omron HEM-9200T, not a separately published exact-model trial. Medaval separately states the equivalence results are unpublished and cannot be verified for its MDR assessment. Maker manual cites an 85-subject EN ISO 81060-2:2014 investigation, excluding pregnant and pre-eclamptic patients, and compliance with 2019+A1:2020+A2:2024. These blood-pressure assessments do not establish diagnostic accuracy of rhythm indicators |
| Features | Coloured risk scale, irregular heartbeat warning, automatic switch-off and incorrect usage message; no cuff fit check or resting indicator | Irregular-heartbeat, body-movement, cuff-wrap and high-blood-pressure indicators. No AFib screening or ECG recording |
| Cuff | ||
| Cuff fits | 22–44 cm | 22–42 cm |
| Cuff | Universal upper-arm cuff supplied | Included HEM-RML31 soft cuff fits 22–42 cm arms; approximately 145 × 594 mm, 750 mm air tube and 170 g. Other compatible cuffs are optional, rather than part of the included cuff range |
| Connections and memory | ||
| Users | 2 | – |
| Readings stored | 60 per user (better) | 30 |
| Bluetooth and app | Yes | Yes |
| Memory | Average of all saved readings and morning/evening averages for the last seven days; date and time stored | Stores up to 30 readings on the monitor |
| App | Bluetooth transfer to the free beurer HealthManager Pro app | Bluetooth Low Energy transfers measurements to OMRON connect and synchronises the monitor date/time; 2.4 GHz band, 2400–2483.5 MHz, GFSK modulation and effective radiated power below 20 dBm |
| Power | Four 1.5 V LR03 AAA batteries | Four AA batteries or optional compatible AC adapter; DC 6 V, 4 W. Approximately 900 measurements with new alkaline batteries and the included cuff, depending on batteries and cuff |
| Design | ||
| Display details | Non-illuminated LCD, 7.6 × 7.0 cm; illuminated control button | Digital LCD; monitor IP21 |
| Size | 13.90 × 9.40 × 4.80 cm (L × W × H) | Approximately 98 × 79 × 138 mm (W × H × L), 260 g without batteries, from the exact HEM-7188T1-LE manual. UK web dimensions are inconsistent with the manual and are not used |
| In the box | Monitor, 22–44 cm cuff, four AAA batteries, storage bag, quick guide and blood pressure pass | Monitor, HEM-RML31 cuff, four AA batteries and instruction manuals 1/2. Exact manual omits a storage case; maker web contents and accessory marketing conflict, and its CM2 cuff wording is inconsistent with the exact manual |
More comparisons
- Beurer BM 54 vs Withings BPM Connect
- Beurer BM 54 vs Beurer BM 85
- Beurer BM 53 vs Beurer BM 54
- Beurer BM 54 vs Beurer BM 59
- Beurer BC 54 vs Beurer BM 54
- A&D Medical UA-611 vs Beurer BM 54
- Beurer BM 54 vs Microlife BP B1 BT
- Beurer BM 54 vs Microlife BP B3 BT
- Beurer BM 54 vs Microlife BP B4 BT
- Beurer BM 48 vs Omron M2+ Connect (HEM-7188T1-LE)
- Beurer BM 59 vs Omron M2+ Connect (HEM-7188T1-LE)
- Beurer BM 54 vs Omron M4 Intelli IT
- Beurer BM 54 vs Rossmax X3 BT