Microlife BP W70 BT vs Omron RS3 Intelli IT

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BP W70 BT compared with RS3 Intelli IT
MicrolifeBP W70 BTOmronRS3 Intelli IT
Overview
Released––
Measuring
TypeWristWrist
Clinically validated–Yes
Irregular heartbeat detectionYesYes
AFib detection or ECGNoNo
MethodAutomatic oscillometric wrist measurement during inflation, Comfort+ pressure control and Quiet Inflation technology. European BP W70 BT; non-Bluetooth W70 is separate hardware. First-sale date not establishedOscillometric wrist measurement; HEM-6161T-E. First on-sale date not established
AccuracyManufacturer's BP3KN1-3B regulatory-filed US draft specifies static pressure accuracy ±3 mmHg and pulse ±5% of readout. European product page gives systolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 beats/minute; cuff-pressure display range 0–299 mmHg is separateMaker's pressure accuracy ±3 mmHg and pulse accuracy ±5%; systolic range 60–260 mmHg, diastolic 40–215 mmHg and pulse 40–180 beats/minute
Clinical validationEuropean maker claims clinical ISO validation but independent exact-model study not recovered; headline omitted. FCC radio authorisation is not clinical validationMedaval attributes STRIDE BP adult self-measurement recommendation to HEM-6181 equivalence, noting that the unpublished assessment cannot be verified against MDR requirements. Exact manual says pregnancy use has not been validated
FeaturesEuropean page describes MAM three-measurement averaging, while Asia-Pacific page describes SmartMAM without a count. US regulatory draft describes three readings with 15-second intervals and an additional reading if needed, despite stray double-measurement wording. Keep regional mode differences explicit. Ordinary IHB, cuff-fit and movement checks; no dedicated AFib screeningCuff-wrap, hypertension and irregular-heartbeat indicators; no advanced wrist-height sensor, movement indicator, averaging, AFib screening or ECG. Web FAQ positioning wording does not establish a sensor
Cuff
Cuff fits13.5–21.5 cm13.5–21.5 cm
CuffEuropean and French maker pages specify 13.5–21.5 cm wrist cuff; US BP3KN1-3B draft specifies 13.5–23 cm. Headline retains European range rather than transferring the larger regional cuffWrist circumference 13.5–21.5 cm
Connections and memory
Users2 (better)1
Readings stored99 per user (better)30
Bluetooth and appYesYes
MemoryFrench maker page confirms two users with 99 stored readings each. US draft describes time-button user selection, last-three-reading average excluding suboptimal cuff fit, oldest-record overwrite and retention after battery replacement; these operating details are regional draft evidence, not a confirmed final European instruction revisionUp to 30 internal readings for one user; stored results transfer to the app. The exact instructions do not describe browsing earlier readings on the monitor
AppBluetooth Low Energy transfer to Microlife Connected Health+ (European maker). Asia-Pacific maker specifies Bluetooth 4.2; exact European version omitted. Regulatory-filed US draft has older pairing instructions, so current app controls may differBluetooth low energy transfers readings to OMRON connect automatically within one hour or manually using the Bluetooth button; device compatibility depends on the app
PowerTwo 1.5 V AAA alkaline batteries (European maker); US regulatory draft claims approximately 300 measurements with new batteries. No internal charging establishedTwo AAA alkaline batteries included; approximately 300 measurements with new alkaline batteries
Design
Display detailsLCD with memory, ON/OFF and time controls; wrist-position guidance and traffic-light scale. US regulatory draft describes blue LED and five height-guidance levels; its IP22 rating is not transferred as a confirmed European ratingDigital LCD; IP22
Size82 × 64 × 13.9 mm. European, French and Asia-Pacific sites state 78 ±5 g including batteries; US regulatory draft states 120 ±5 g including batteries. Conflicting regional mass claims retained rather than treated as a settled measurementApproximately 84 × 62 × 21 mm (W × H × L, excluding cuff); 2022 manual gives 85 g without batteries, versus web table's rounded 90 g
In the boxUS regulatory-filed draft lists instructions, storage box, blood-pressure diary and two AAA cells; European package contents not independently confirmedMonitor, two AAA alkaline batteries, storage case, instruction manual and setup instructions

More comparisons