Beurer BM 23 vs Microlife BP N2 Easy

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BM 23 compared with BP N2 Easy
BeurerBM 23MicrolifeBP N2 Easy
Overview
Released––
Measuring
TypeUpper armUpper arm
Irregular heartbeat detectionYesYes
AFib detection or ECGNoNo
MethodAutomatic oscillometric upper-arm measurement during deflation. Exact UK item 60493, EAN 4211125604936; manual model BM 23, technical type BM 25. Range tile and SEO title use BM 25, but active product heading, photos, datasheet and instructions identify BM 23. First-sale date unresolvedSemi-automatic oscillometric upper-arm measurement. User inflates cuff with supplied pump ball; electronic measurement follows during deflation. Quick-acting valve releases remaining cuff air. No automatic triple-measurement mode. First actual retail date unresolved
AccuracyManual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse; systolic range 50–280 mmHg, diastolic 30–200 mmHg and pulse 40–199 beats/minute. Maximum cuff pressure 300 mmHg is separateExact April 2021 guide gives blood pressure 20–280 mmHg and pulse 40–200 bpm; global support instead lists systolic 60–255 and diastolic 40–200 mmHg. Asia-Pacific product page gives pulse upper limit 199 bpm. Static pressure accuracy ±3 mmHg and pulse accuracy ±5%; cuff-pressure display 0–299 mmHg is separate
Clinical validationCurrent manual claims clinical testing to ISO 81060-2. Medaval lists a Stride BP derivative recommendation based on equivalence to BM 28 via Globalcare GCE603, but says unpublished equivalence results cannot be verified to MDR requirements. Independently established exact-device validation remains unresolved; headline flag omittedMaker guide claims shared measuring technology with BP 3BTO-H tested under ESH protocol and broader special-population suitability. Medaval reports no peer-reviewed exact-device validation evidence; bounded STRIDE search did not recover exact N2 Easy acceptance. Independent exact validation remains unresolved, so headline validation is omitted
FeaturesIrregular-heartbeat warning, coloured classification scale, electronic cuff-position check, automatic switch-off and battery indicator. No ECG, dedicated AFib screening, resting indicator or voice outputGeneric IHB, termed PAD on Asia-Pacific product page; not AFib-specific screening or ECG. Hypertension warning, pump-up and air-discharge indicators, low-battery warning and automatic switch-off after approximately one minute. No date/time controls
Cuff
Cuff fits22–42 cm22–32 cm
CuffSupplied universal cuff for 22–42 cm upper-arm circumference, replacement item 110.031; conventional hook-and-loop cuff rather than easyClip or easyLockAsia-Pacific bundle supplies conical M 22–32 cm cuff. April 2021 guide lists optional S 17–22 and L 32–42 cm; L cuff requires corresponding black deflation valve. Optional S cuff needs original cuff connector. Pump and cuff can use either of the two labelled pneumatic sockets; these are not dual-arm electronic measurement channels
Connections and memory
Users–1
Readings stored–1
Bluetooth and appNoNo
MemoryNo stored reading history, user profiles, date/time or calculated averages established by the product table or complete device operation section. Manual's isolated statement about resetting date/time and retaining saved measurements conflicts with those sections; not treated as an implemented featureOne shared stored reading, replaced by the next result. Hold ON/OFF while switched off to recall. No multi-reading history, user selector, timestamps or averaging function in exact guide
AppNo Bluetooth connection from the monitor. Scan & Save captures the display with a smartphone camera into beurer HealthManager Pro rather than automatically transmitting measurementsNo Bluetooth, USB or PC-transfer function in exact controls
PowerFour 1.5 V LR03 AAA batteries included; manual approximately 300 measurements depending on blood pressure and inflation pressure. USB-C input supports optional 5 V DC / 1 A mains operation, not battery charging or a demonstrated data connection. Manual lists EU adapter 072.78 and UK adapter 072.79; remove batteries for mains operationTwo 1.5 V AA alkaline cells; approximately 1500 measurements with new batteries. NiMH cells require external charging. No mains-adaptor socket or operating instructions in exact labelled controls; generic adaptor wording in guarantee exclusions does not establish mains support
Design
Display detailsNon-illuminated LCD with black digits, maker 4.4 × 5.3 cm; physical START/STOP button. Manual IP21LCD showing systolic, diastolic, pulse, IHB, stored-value, battery and inflation/deflation indicators. One ON/OFF button and two pneumatic sockets. IP20
Size120 × 90 × 40 mm (L × W × H). Manual and datasheet approximately 318 g with cuff and without batteries; current product table's 362 g excluding packaging has unspecified weighing conditions136 × 82 × 50 mm; 205 g including batteries in exact guide and global support
In the boxMonitor, universal upper-arm cuff, four AAA batteries and instructions according to the datasheet and manual. Product page additionally lists a blood pressure record; mains adapter sold separatelyAsia-Pacific page lists monitor, two AA alkaline cells, pump ball, blood-pressure diary and M 22–32 cm cuff. Regional bundle scope retained; accessories are not inferred from other Microlife monitors