Microlife BP A6 BT vs Omron M7 Intelli IT AFib

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BP A6 BT compared with M7 Intelli IT AFib
MicrolifeBP A6 BTOmronM7 Intelli IT AFib
Overview
Released–September 2024
Measuring
TypeUpper armUpper arm
Clinically validatedYesYes
Irregular heartbeat detectionYesYes
AFib detection or ECGYesYes
MethodOscillometric upper-arm measurement during deflation with Gentle+ control. Single or AFIB/MAM triple mode selected by side switch; three measurements separated by 15 seconds produce one analysed result. Questionable measurement triggers a fourth; individual triple-mode results are not displayed. Actual first retail sale unresolvedAutomatic oscillometric upper-arm measurement; model HEM-7380T1-EBK
AccuracyCurrent E-V11 MDR guide gives systolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 bpm; global support lists pulse 40–200. Older S-V10 guide and French maker page give blood pressure 20–280 mmHg and pulse 40–200. Static accuracy ±3 mmHg and pulse accuracy ±5%; cuff-pressure display 0–299 mmHg is separatePressure ±3 mmHg; pulse ±5% of the displayed reading
Clinical validationSTRIDE BP accepts exact A6 BT for adult and diabetic home use by equivalence to BP A100, with referenced 2009, 2018 and 2021 studies. This does not establish separate published A6 BT trials for every special population mentioned by the maker. Blood-pressure validation and AFib screening are separate claimsSTRIDE BP lists HEM-7380T1-EBK as an approved equivalent of Omron BP760N, for adults, type 2 diabetes and pregnancy including preeclampsia
FeaturesAFib pulse-irregularity screening active only in AFIB/MAM triple mode, not single measurement. Does not record an ECG or provide an ECG-confirmed diagnosis. Cuff-fit and arm-movement indications, LED pressure classification scale, low-battery warning, automatic switch-off and lock switch to prevent accidental activationScreens for possible AFib from pulse waves during routine readings; does not record an ECG. Irregular heartbeat and body movement indicators, cuff wrap guide and TruRead three-reading averaging
Cuff
Cuff fits22–42 cm22–42 cm
CuffCurrent MDR guide supplies Microlife Soft Cuff M-L 22–42 cm. Optional soft cuffs S 17–22, M 22–32, L 32–42 and L-XL 32–52 cm are not all supplied. STRIDE lists validated 22–42 and 32–52 cm sizesHEM-FL31 Intelli Wrap Cuff, 22 to 42 cm; 750 mm air tube
Connections and memory
Users12 (better)
Readings stored99100 per user (better)
Bluetooth and appYesYes
MemoryOne history of the last 99 dated readings, with no user selector in exact controls. First recall shows average of all stored values; older guide calls it 28A without defining a 28-day interval. Forward/backward buttons navigate history, oldest overwritten when full. Complete history can be cleared, individual saved values cannot; newly displayed reading can be excluded. Readings survive battery replacement while clock requires reset. Isolated manual wording per user does not establish two separate histories100 readings per user for two users, plus guest mode; morning, evening and weekly averages
AppBluetooth 4.0 synchronisation with Microlife Connected Health+; connection active after measurement and can be started with forward button in standby. USB transfer to Blood Pressure Analyzer+ PC software. Both manuals specify Mini-B five-pin connector in PC-link section, while current package list calls its cable Micro USB; connector wording conflict retained. French page keeps older app OS requirements; current supported phone versions should be checked with the appOMRON connect over Bluetooth Low Energy
PowerFour 1.5 V AAA alkaline cells; approximately 400 measurements with new batteries, not a separate triple-cycle count. Optional Microlife DC 6 V, 600 mA adapter; French bundle list includes it despite technical table calling it optional. NiMH cells supported with external charging, not charged in monitorFour AA batteries included; optional AC adapter. Approximately 700 measurements with new alkaline batteries (fewer when using TruRead)
Design
Display detailsLCD with date/time and AFIB/MAM indicators; START/STOP, memory, time and forward/backward controls. Separate lock and single/triple switches, cuff, USB and DC sockets. IP20Digital LCD showing blood pressure and pulse, with previous-reading comparison
SizeCurrent E-V11 MDR guide, French product and global support agree on 160 × 80 × 32 mm and 354 g including batteries. Older exact S-V10 guide incorrectly gives 150 × 100 × 50 mm and 580 g; conflict retained rather than transferring that table191 × 85 × 117 mm (W × H × D); monitor 440 g without batteries, cuff approximately 163 g
In the boxCurrent MDR guide lists soft M-L cuff, cuff holder, cuff-size marker, USB cable, storage bag, diary, instructions and four AAA cells, with adapter optional. French regional list includes adapter; regional bundle differences remain qualifiedMonitor, Intelli Wrap Cuff, four AA batteries, instruction manuals, storage case and blood pressure pass