Microlife BP A6 BT vs Omron M4 Connect AFib

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

BP A6 BT compared with M4 Connect AFib
MicrolifeBP A6 BTOmronM4 Connect AFib
Overview
Released––
Measuring
TypeUpper armUpper arm
Clinically validatedYesYes
Irregular heartbeat detectionYesYes
AFib detection or ECGYesYes
MethodOscillometric upper-arm measurement during deflation with Gentle+ control. Single or AFIB/MAM triple mode selected by side switch; three measurements separated by 15 seconds produce one analysed result. Questionable measurement triggers a fourth; individual triple-mode results are not displayed. Actual first retail sale unresolvedAutomatic oscillometric upper-arm measurement; HEM-7196T1-FLE. First retail release date has not been established; the manual's December 2024 revision is not an on-sale date
AccuracyCurrent E-V11 MDR guide gives systolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 bpm; global support lists pulse 40–200. Older S-V10 guide and French maker page give blood pressure 20–280 mmHg and pulse 40–200. Static accuracy ±3 mmHg and pulse accuracy ±5%; cuff-pressure display 0–299 mmHg is separateMaker's pressure accuracy ±3 mmHg and pulse accuracy ±5% of displayed reading; systolic range 60–260 mmHg, diastolic 40–215 mmHg and pulse 40–180 beats/minute
Clinical validationSTRIDE BP accepts exact A6 BT for adult and diabetic home use by equivalence to BP A100, with referenced 2009, 2018 and 2021 studies. This does not establish separate published A6 BT trials for every special population mentioned by the maker. Blood-pressure validation and AFib screening are separate claimsSTRIDE BP approves this exact hardware code as equivalent to Omron BP760N (HEM-7320-Z), with adults, diabetes and pregnancy/preeclampsia listed; equivalent-device approval is distinct from a direct test of this model or validation of its AFib algorithm
FeaturesAFib pulse-irregularity screening active only in AFIB/MAM triple mode, not single measurement. Does not record an ECG or provide an ECG-confirmed diagnosis. Cuff-fit and arm-movement indications, LED pressure classification scale, low-battery warning, automatic switch-off and lock switch to prevent accidental activationPulse-wave screening for possible AFib, rather than an ECG or diagnosis; irregular heartbeat, body movement and cuff-wrap indicators. The exact manual describes three consecutive readings at 30-second intervals with their average; this conflicts with the UK page's averaging-function No table entry
Cuff
Cuff fits22–42 cm22–42 cm
CuffCurrent MDR guide supplies Microlife Soft Cuff M-L 22–42 cm. Optional soft cuffs S 17–22, M 22–32, L 32–42 and L-XL 32–52 cm are not all supplied. STRIDE lists validated 22–42 and 32–52 cm sizesHEM-FL31 preformed Intelli Wrap Cuff; approximately 145 × 532 mm, 750 mm air tube and 163 g
Connections and memory
Users12 (better)
Readings stored99 (better)60 per user
Bluetooth and appYesYes
MemoryOne history of the last 99 dated readings, with no user selector in exact controls. First recall shows average of all stored values; older guide calls it 28A without defining a 28-day interval. Forward/backward buttons navigate history, oldest overwritten when full. Complete history can be cleared, individual saved values cannot; newly displayed reading can be excluded. Readings survive battery replacement while clock requires reset. Isolated manual wording per user does not establish two separate historiesUp to 60 readings per user for two users; guest mode takes one measurement without saving it
AppBluetooth 4.0 synchronisation with Microlife Connected Health+; connection active after measurement and can be started with forward button in standby. USB transfer to Blood Pressure Analyzer+ PC software. Both manuals specify Mini-B five-pin connector in PC-link section, while current package list calls its cable Micro USB; connector wording conflict retained. French page keeps older app OS requirements; current supported phone versions should be checked with the appBluetooth Low Energy sync with OMRON connect; readings can be transferred to a compatible smart device
PowerFour 1.5 V AAA alkaline cells; approximately 400 measurements with new batteries, not a separate triple-cycle count. Optional Microlife DC 6 V, 600 mA adapter; French bundle list includes it despite technical table calling it optional. NiMH cells supported with external charging, not charged in monitorFour AA batteries included; optional HHP-CM01, HHP-AM01 or HHP-BFH01 AC adapter. Approximately 900 measurements with new alkaline batteries and the included cuff, depending on batteries and cuff
Design
Display detailsLCD with date/time and AFIB/MAM indicators; START/STOP, memory, time and forward/backward controls. Separate lock and single/triple switches, cuff, USB and DC sockets. IP20Digital LCD; monitor IP21
SizeCurrent E-V11 MDR guide, French product and global support agree on 160 × 80 × 32 mm and 354 g including batteries. Older exact S-V10 guide incorrectly gives 150 × 100 × 50 mm and 580 g; conflict retained rather than transferring that tableApproximately 165 × 70 × 98 mm (W × H × L); monitor approximately 350 g without batteries
In the boxCurrent MDR guide lists soft M-L cuff, cuff holder, cuff-size marker, USB cable, storage bag, diary, instructions and four AAA cells, with adapter optional. French regional list includes adapter; regional bundle differences remain qualifiedMonitor, HEM-FL31 cuff, four AA batteries, instruction manuals 1/2, storage case and blood pressure pass