A&D Medical UA-611 vs Microlife BP 3AG1
Better figure, where one is clearly better
| A&D MedicalUA-611 | MicrolifeBP 3AG1 | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Upper arm | Upper arm |
| Irregular heartbeat detection | Yes | Yes |
| AFib detection or ECG | – | No |
| Method | Automatic oscillometric measurement with fuzzy-logic-controlled inflation; designed for adult home use; first on-sale date not established | Oscillometric upper-arm measurement during deflation. One automatic measurement started with ON/OFF button; no MAM triple-mode selector in exact guide. Actual first retail sale unresolved |
| Accuracy | Maker specification ±3 mmHg pressure and ±5% pulse; pressure range 0–299 mmHg, systolic 60–279 mmHg, diastolic 40–200 mmHg and pulse 40–180 beats/min | Exact March 2023 guide gives blood pressure 20–280 mmHg and pulse 40–200 bpm; current global support instead lists systolic 60–255 and diastolic 40–200 mmHg. Static pressure accuracy ±3 mmHg and pulse accuracy ±5%; cuff-pressure display 0–299 mmHg is separate |
| Clinical validation | A&D UK claims ESH 2010 validation; Medaval identifies the UK BIHS recommendation as manufacturer-declared equivalence to the UA-651 and says the equivalence study is unpublished. The US Validate BP listing covers UA-611 under ANSI/AAMI/ISO 81060-2:2018 with different cuff options; this does not establish direct validation of the supplied UK cuff configuration | Exact guide claims shared measurement technology with BP 3BTO-A tested under BIHS protocol. Independent exact BP 3AG1 validation was not recovered in the bounded registry search, so headline validation is omitted. Broad special-population wording in maker guide is not promoted to independently established exact-model validation |
| Features | Irregular heartbeat indicator, pressure progress bar, WHO classification indicator, stored-reading average and automatic power-off; irregular heartbeat indication is not a specific AFib diagnosis | Generic irregular-heartbeat indication, not AFib-specific screening or ECG recording. Low-battery warning, pulse indicator and automatic switch-off after approximately one minute. No date/time controls or automatic averaging mode in exact guide |
| Cuff | ||
| Cuff fits | 22–32 cm | – |
| Cuff | Supplied latex-free and metal-free SlimFit cuff fits 22–32 cm; optional large cuff fits 31–45 cm, sold separately; UK manual says this model is not designed for a small cuff | Supplied cuff size not established in recovered global support and guide. Optional S 17–22, M 22–32, M-L 22–42 and L 32–42 cm cuffs and optional pre-shaped cuffs listed; these are not all supplied. Optional small cuff needs the connector from original supplied cuff |
| Connections and memory | ||
| Users | – | 1 |
| Readings stored | 30 | 30 |
| Bluetooth and app | No | No |
| Memory | Last 30 blood pressure and pulse measurements, with stored-reading average; no separate user memories listed | One history of the last 30 readings, with no user selector or timestamps. First recall displays count then latest reading; repeated memory presses show earlier values. Oldest overwritten when full. Complete history can be cleared, individual saved values cannot; newly displayed reading can be excluded |
| App | No Bluetooth; readings can be recorded manually | No Bluetooth, USB socket or PC-transfer function in exact guide and labelled controls |
| Power | Four AA 1.5 V batteries; maker claims approximately 700 measurements with LR6 alkaline batteries or 200 with R6P manganese batteries at 180 mmHg and 23°C. UA-611 has no DC socket; the optional adapter in the shared manual applies only to UA-651 | Four 1.5 V AA alkaline cells; approximately 920 measurements with new batteries. Optional Microlife DC 6 V, 600 mA adapter stops battery-current use while connected. NiMH cells supported with external charging, not charged inside monitor |
| Design | ||
| Display details | Systolic pressure, diastolic pressure and pulse shown with battery, irregular heartbeat and pressure/classification indicators | LCD with systolic, diastolic, pulse, irregular-heartbeat, battery and stored-value indications. ON/OFF and memory buttons, cuff and DC sockets. IP20 |
| Size | Approximately 96 × 68 × 130 mm (width × height × depth); approximately 230 g excluding batteries | 140 × 120 × 70 mm; 400 g including batteries in exact guide and global maker support |
| In the box | UA-611 monitor, 22–32 cm SlimFit cuff, instruction manual and four AA batteries in the UK pack | Exact global supplied-accessory list not recovered; optional cuff sizes and mains adapter are not treated as bundled. Guide filename contains 3AG1-30 and W identifiers while its exact cover and controls identify BP 3AG1; no cosmetic duplicate record created |
More comparisons
- A&D Medical UA-611 vs A&D Medical UA-611Plus
- A&D Medical UA-611 vs Beurer BM 54
- A&D Medical UA-611 vs Rossmax AU941f
- A&D Medical UA-611 vs Rossmax AX356f
- A&D Medical UA-611 vs Beurer BM 23
- Beurer BM 23 vs Microlife BP 3AG1
- A&D Medical UA-611 vs Beurer BM 25
- Beurer BM 25 vs Microlife BP 3AG1
- A&D Medical UA-611 vs Beurer BM 27
- Beurer BM 27 vs Microlife BP 3AG1
- A&D Medical UA-611 vs Beurer BM 28
- Beurer BM 28 vs Microlife BP 3AG1
- A&D Medical UA-611 vs Beurer BM 35
- Beurer BM 35 vs Microlife BP 3AG1
- A&D Medical UA-611 vs Beurer BM 38
- Beurer BM 38 vs Microlife BP 3AG1