Beurer BM 25 vs Microlife BP A3 PC
Better figure, where one is clearly better
| BeurerBM 25 | MicrolifeBP A3 PC | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Upper arm | Upper arm |
| Clinically validated | – | Yes |
| Irregular heartbeat detection | Yes | Yes |
| AFib detection or ECG | – | No |
| Method | Automatic oscillometric upper-arm measurement during deflation; first on-sale date not established | Oscillometric upper-arm measurement during deflation. Side switch selects a single measurement or MAM mode, which analyses three readings with 15-second pauses into one displayed result. Individual MAM results are not displayed; a fourth measurement is taken if one is questionable. First actual on-sale date unresolved |
| Accuracy | Maker manual specifies display accuracy of ±3 mmHg for systolic and diastolic pressure and ±5% of the displayed pulse value; this specification is distinct from clinical validation | Exact April 2020 English/Greek guide gives blood pressure 20–280 mmHg, pulse 40–200 bpm, static accuracy ±3 mmHg and pulse accuracy ±5%. Current global support instead separates systolic 60–255 and diastolic 40–200 mmHg; range difference remains explicit. Cuff-pressure display 0–299 mmHg is separate |
| Clinical validation | Beurer states clinical testing to ISO 81060-2 in the manual; exact independent registry confirmation was not established | STRIDE BP lists an exact-model diabetes study using ANSI/AAMI/ISO 2013, published in 2018, and adult acceptance by equivalence to BP A100. Published Beime study abstract reports 85 diabetic participants aged 56–88 and passing both protocol criteria; this does not establish every population in broad maker claims. Registry lists 22–42 and 32–52 cm validated cuffs |
| Features | Single start/stop button, cuff fit check, irregular heartbeat warning, classification scale and automatic switch-off; maker lists no resting indicator or calculated averages | Pulse Arrhythmia Detection indicates pulse irregularities, not a separate AFib algorithm or ECG. Cuff-fit and arm-movement indicators, coloured classification scale and low-battery warning; automatic switch-off after approximately one minute |
| Cuff | ||
| Cuff fits | 22–42 cm | – |
| Cuff | Supplied upper-arm cuff | Exact guide lists S 17–22, M 22–32, M-L 22–42, L 32–42 and L-XL 32–52 cm cuffs and optional preformed Easy cuffs. Recovered support record does not establish the supplied size, so headline cuff range is omitted. Guide describes removable bladder and hand-washable cuff cover; do not wash bladder |
| Connections and memory | ||
| Users | – | 2 |
| Readings stored | – | 99 per user |
| Bluetooth and app | No | No |
| Memory | – | User switch selects one of two separate histories, each storing 99 dated readings. First recall shows average of all stored values, followed by individual history. Oldest overwritten when full. All values for selected user can be cleared, but individual stored values cannot; a newly displayed result can be excluded before storage. History survives battery replacement while clock requires resetting |
| App | Scan & Save in beurer HealthManager Pro uses a smartphone camera to record the display; no Bluetooth connection | USB PC link to Microlife Blood Pressure Analyzer+ using Mini-B 5-pin connector. Computer controls device during connection. No Bluetooth or mobile-app transfer. Current software compatibility must be checked separately rather than inferred from old CDs |
| Power | Four 1.5 V LR03 AAA batteries or optional DC 5 V, 1 A mains adaptor through USB-C; USB-C is for mains operation, not charging the disposable batteries | Four 1.5 V AA alkaline cells; approximately 920 measurements with new batteries, without a separate MAM-cycle count. Optional Microlife DC 6 V, 600 mA adapter stops battery-current use when connected. NiMH cells supported but require external charging |
| Design | ||
| Display details | Non-illuminated LCD, 4.4 × 5.3 cm, with black digits; maker lists no date/time display | LCD with date/time and user indicator, ON/OFF, memory and time buttons, separate user and 1/3 MAM switches. USB and DC adapter sockets. IP20 |
| Size | 12.00 × 9.00 × 4.00 cm (L × W × H) | 143 × 85 × 58 mm; 368 g including batteries in exact guide and global support |
| In the box | Monitor, 22–42 cm cuff and four AAA batteries | Exact guide allows regional packages without cable or download voucher; complete original supplied bundle unresolved. Verified global archived specification/support record serves both resource types. English guide includes Greek regional guarantee wording; this is not a worldwide warranty claim |
More comparisons
- Beurer BM 25 vs Beurer BM 85
- Beurer BM 25 vs Beurer BM 53
- Beurer BM 25 vs Rossmax AU941f
- Beurer BM 25 vs Rossmax AX356f
- A&D Medical UA-611 vs Beurer BM 25
- Beurer BM 25 vs Microlife BP 3AG1
- Beurer BM 25 vs Microlife BP 3AS1-2
- Beurer BM 25 vs Microlife BP A1 Easy
- Beurer BM 25 vs Microlife BP A2 Basic
- Beurer BM 25 vs Microlife BP A2 Classic
- Beurer BM 25 vs Microlife BP A2 Easy
- A&D Medical UA-611 vs Microlife BP A3 PC
- Beurer BM 23 vs Microlife BP A3 PC
- Beurer BM 25 vs Microlife BP A3 Plus
- Beurer BM 25 vs Microlife BP A3L Comfort
- Beurer BM 25 vs Microlife BP AG1-10