BeurerBM 64OmronEvolv
Overview
Released–January 2017
Measuring
TypeUpper armUpper arm
Clinically validated–Yes
Irregular heartbeat detectionYesYes
AFib detection or ECGYesYes
MethodAutomatic oscillometric upper-arm measurement during inflation. UK item 65400, EAN 4211125654009; manual type BM 64. Announced on 12 November 2023 with planned German availability from March 2024; actual first-sale date unresolved–
AccuracyManual display accuracy ±3 mmHg for systolic and diastolic pressure and ±5% for pulse. Systolic range 50–280 mmHg, diastolic 30–200 mmHg and pulse 40–199 beats/minute; maximum cuff pressure 300 mmHg is separate–
Clinical validationCurrent manual claims clinical testing to ISO 81060-2; independent validation for this exact model not established. The shared launch release's pregnancy-validation claim belongs to BM 53, not BM 64Listed as validated for home use by STRIDE BP (HEM-7600T-E); Omron lists clinical and pregnancy validation
FeaturesAdvanced arrhythmia detection flags possible atrial fibrillation, without recording an ECG. Coloured classification scale, cuff-position check, HSD resting indicator and low-battery indication. Maker retains triple-measurement mode; May 2024 instructions describe three measurements with 30-second waits and a final average, while September 2026 instructions omit that operation sectionIntellisense inflation; cuff wrap guide; body movement detection; irregular heartbeat and AFib detection
Cuff
Cuff fits22–42 cm22–42 cm
CuffSupplied premium cuff fits upper-arm circumferences of 22–42 cm; conventional tube connection, not easyClip or easyLockIntelli Wrap Cuff built into the monitor, 22 to 42 cm
Connections and memory
Users2–
Readings stored120 per user (better)100
Bluetooth and appYesYes
MemoryTwo separately selected users with 120 readings each; oldest reading replaced when full. Recall individual readings and average of all stored readings, plus morning (05:00–09:00) and evening (18:00–20:00) averages over the last seven days. Date/time and user-selection slider explicitly described in the manual100 readings
AppBluetooth transfer to beurer HealthManager Pro with automatic transfer configurable on the monitor; six-digit pairing PIN. Manual publishes 2400–2483 MHz band and maximum 5 dBm transmission power, without a Bluetooth versionOMRON connect, over Bluetooth
PowerFour included 1.5 V LR6 AA batteries; approximately 300 measurements depending on blood pressure, inflation pressure and Bluetooth connections. Optional USB-C mains supply 5 V DC / 1 A; remove batteries for mains use. Included USB-C cable does not charge batteriesBatteries (set included)
Design
Display detailsIlluminated inverse XL LCD with white digits on black; four sensor touch buttons and a central START/STOP control. Manual IP20–
Size143 × 105 × 60 mm (L × W × H). Manual and datasheet approximately 450 g with cuff and without batteries; current UK website lists 646 g excluding packaging without specifying cuff/battery weighing conditions85 × 20 × 120 mm; 240 g
In the boxMonitor, 22–42 cm premium cuff, four AA batteries, USB-C cable, storage pouch and instructions; mains adapter optionalEVOLV blood pressure monitor, instruction manual, storage case, battery set, blood pressure pass, OMRON connect set-up instructions