A&D Medical UA-611 vs Microlife BP A2 Basic

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

UA-611 compared with BP A2 Basic
A&D MedicalUA-611MicrolifeBP A2 Basic
Overview
Released––
Measuring
TypeUpper armUpper arm
Clinically validated–Yes
Irregular heartbeat detectionYesYes
AFib detection or ECG–No
MethodAutomatic oscillometric measurement with fuzzy-logic-controlled inflation; designed for adult home use; first on-sale date not establishedAutomatic oscillometric upper-arm measurement during deflation, marketed as Gentle+ pressure and speed control. Exact operation describes pumping followed by gradual pressure reduction; no MAM sequence established despite copied MAM-button wording in some instructions. First on-sale date unresolved
AccuracyMaker specification ±3 mmHg pressure and ±5% pulse; pressure range 0–299 mmHg, systolic 60–279 mmHg, diastolic 40–200 mmHg and pulse 40–180 beats/minJuly 2025 E-V10 manual gives systolic 60–255 mmHg, diastolic 40–200 mmHg and pulse 40–199 bpm; static pressure ±3 mmHg and pulse ±5%. UK product/support instead give blood pressure 20–280 mmHg and pulse 40–200 bpm; September 2023 S-V11 manual says pulse 20–200 despite its error table's 40 bpm threshold. Conflicting ranges qualified; cuff-pressure display 0–299 mmHg is separate
Clinical validationA&D UK claims ESH 2010 validation; Medaval identifies the UK BIHS recommendation as manufacturer-declared equivalence to the UA-651 and says the equivalence study is unpublished. The US Validate BP listing covers UA-611 under ANSI/AAMI/ISO 81060-2:2018 with different cuff options; this does not establish direct validation of the supplied UK cuff configurationSTRIDE BP lists A2 Basic as an equivalent device to BP A100, for adults and diabetes with 22–42 and 32–52 cm cuffs. Headline indicates registry acceptance by equivalence, not a separate exact-model trial. Maker manual instead cites BP 3BT0-A technology; Medaval notes that the equivalence evidence does not establish MDR-compliant accuracy validation
FeaturesIrregular heartbeat indicator, pressure progress bar, WHO classification indicator, stored-reading average and automatic power-off; irregular heartbeat indication is not a specific AFib diagnosisOrdinary IHB indication, traffic-light classification, cuff check, movement warning, low-battery indicator and automatic switch-off after approximately one minute. No dedicated AFib screening or app
Cuff
Cuff fits22–32 cm22–42 cm
CuffSupplied latex-free and metal-free SlimFit cuff fits 22–32 cm; optional large cuff fits 31–45 cm, sold separately; UK manual says this model is not designed for a small cuffSupplied conical M-L cuff fits 22–42 cm in UK maker product page. Manual lists optional S 17–22, M 22–32, L 32–42 and L-XL 32–52 cm; do not treat all optional sizes as supplied
Connections and memory
Readings stored3030
Bluetooth and appNoNo
MemoryLast 30 blood pressure and pulse measurements, with stored-reading average; no separate user memories listedStores last 30 dated readings and first displays their average on recall, then individual results. Oldest overwritten when full. History retained on cell replacement while clock resets. Newly displayed reading can be excluded; stored history clears together. No user-selector operation established; manuals contain copied two-user wording, so headline user count remains omitted
AppNo Bluetooth; readings can be recorded manually–
PowerFour AA 1.5 V batteries; maker claims approximately 700 measurements with LR6 alkaline batteries or 200 with R6P manganese batteries at 180 mmHg and 23°C. UA-611 has no DC socket; the optional adapter in the shared manual applies only to UA-651Four 1.5 V AA alkaline batteries, approximately 920 measurements with new cells in July 2025 manual. Optional Microlife DC 6 V, 600 mA mains adapter uses dedicated socket without drawing battery current. NiMH rechargeable cells supported but charged externally; no internal charging
Design
Display detailsSystolic pressure, diastolic pressure and pulse shown with battery, irregular heartbeat and pressure/classification indicatorsLCD with ON/OFF, memory and time controls, cuff socket and adapter socket; IP20 in July 2025 manual
SizeApproximately 96 × 68 × 130 mm (width × height × depth); approximately 230 g excluding batteriesUK product/support and July 2025 E-V10 manual agree on 135.5 × 82 × 57 mm and 340 g including batteries. September 2023 S-V11 instead says 157.5 × 105 × 61.5 mm and 415 g; those conflicting older figures are not assigned to the headline
In the boxUA-611 monitor, 22–32 cm SlimFit cuff, instruction manual and four AA batteries in the UK packMonitor, textile storage pouch, four AA alkaline cells, blood-pressure diary and M-L 22–42 cm cuff listed by UK maker page; mains adapter optional