A&D Medical UA-611 vs Microlife BP AG1-20
Better figure, where one is clearly better
| A&D MedicalUA-611 | MicrolifeBP AG1-20 | |
|---|---|---|
| Overview | ||
| Released | – | – |
| Measuring | ||
| Type | Upper arm | Upper arm |
| Irregular heartbeat detection | Yes | No |
| AFib detection or ECG | – | No |
| Method | Automatic oscillometric measurement with fuzzy-logic-controlled inflation; designed for adult home use; first on-sale date not established | Fully manual auscultatory upper-arm measurement using the supplied stethoscope and aneroid pressure gauge. Hand inflation and screw-controlled deflation, with readings interpreted by the operator. Actual first retail date unresolved |
| Accuracy | Maker specification ±3 mmHg pressure and ±5% pulse; pressure range 0–299 mmHg, systolic 60–279 mmHg, diastolic 40–200 mmHg and pulse 40–180 beats/min | March 2021 exact guide gives 0–300 mmHg range and 2 mmHg scale divisions. Static accuracy within ±3 mmHg at 18–33°C and ±6 mmHg at 34–46°C; hysteresis error 0–4 mmHg and air leakage under ±4 mmHg/min. Maker recommends an accuracy test every two years or after mechanical impact |
| Clinical validation | A&D UK claims ESH 2010 validation; Medaval identifies the UK BIHS recommendation as manufacturer-declared equivalence to the UA-651 and says the equivalence study is unpublished. The US Validate BP listing covers UA-611 under ANSI/AAMI/ISO 81060-2:2018 with different cuff options; this does not establish direct validation of the supplied UK cuff configuration | Guide cites EN ISO 81060-1 and ANSI/AAMI SP09. These standards are separate from independent clinical validation; exact independent validation unresolved and headline omitted |
| Features | Irregular heartbeat indicator, pressure progress bar, WHO classification indicator, stored-reading average and automatic power-off; irregular heartbeat indication is not a specific AFib diagnosis | Mechanical dial, separate pump ball and adjustable deflation valve. No automatic pressure or pulse calculation, irregular-heartbeat screening, AFib detection or ECG |
| Cuff | ||
| Cuff fits | 22–32 cm | 22–32 cm |
| Cuff | Supplied latex-free and metal-free SlimFit cuff fits 22–32 cm; optional large cuff fits 31–45 cm, sold separately; UK manual says this model is not designed for a small cuff | Supplied soft M 22–32 cm cuff established by exact current guide and Asia-Pacific product page. Other regional bundles are not inferred from retailer variants |
| Connections and memory | ||
| Readings stored | 30 | – |
| Bluetooth and app | No | No |
| Memory | Last 30 blood pressure and pulse measurements, with stored-reading average; no separate user memories listed | No electronic storage; readings, date and time recorded separately by the operator |
| App | No Bluetooth; readings can be recorded manually | No Bluetooth, USB or phone app |
| Power | Four AA 1.5 V batteries; maker claims approximately 700 measurements with LR6 alkaline batteries or 200 with R6P manganese batteries at 180 mmHg and 23°C. UA-611 has no DC socket; the optional adapter in the shared manual applies only to UA-651 | Mechanical operation without batteries or mains power |
| Design | ||
| Display details | Systolic pressure, diastolic pressure and pulse shown with battery, irregular heartbeat and pressure/classification indicators | Analogue manometer with pointer; no LCD |
| Size | Approximately 96 × 68 × 130 mm (width × height × depth); approximately 230 g excluding batteries | Product and support pages list 175 × 70 × 103 mm and 450 g. March 2021 exact guide instead lists 185 × 75 × 110 mm (±1 mm) and weight up to 540 g. Conditions and the reason for the discrepancy are not established |
| In the box | UA-611 monitor, 22–32 cm SlimFit cuff, instruction manual and four AA batteries in the UK pack | Current guide lists M cuff, pump ball, valve, stethoscope and soft bag; manometer identified in controls |
More comparisons
- A&D Medical UA-611 vs A&D Medical UA-611Plus
- A&D Medical UA-611 vs Beurer BM 54
- A&D Medical UA-611 vs Rossmax AU941f
- A&D Medical UA-611 vs Rossmax AX356f
- A&D Medical UA-611 vs Beurer BM 23
- A&D Medical UA-611 vs Beurer BM 25
- A&D Medical UA-611 vs Beurer BM 27
- A&D Medical UA-611 vs Beurer BM 28
- A&D Medical UA-611 vs Beurer BM 35
- A&D Medical UA-611 vs Beurer BM 38
- A&D Medical UA-611 vs Beurer BM 51 easyClip
- A&D Medical UA-611 vs Beurer BM 58
- A&D Medical UA-611 vs Microlife BP 3AG1
- A&D Medical UA-611 vs Microlife BP 3AS1-2
- A&D Medical UA-611 vs Microlife BP A1 Easy
- A&D Medical UA-611 vs Microlife BP A2 Basic