A&D Medical UA-611 vs Microlife BP N2 Easy

Every spec side by side, from each maker's published figures.

Better figure, where one is clearly better

UA-611 compared with BP N2 Easy
A&D MedicalUA-611MicrolifeBP N2 Easy
Overview
Released––
Measuring
TypeUpper armUpper arm
Irregular heartbeat detectionYesYes
AFib detection or ECG–No
MethodAutomatic oscillometric measurement with fuzzy-logic-controlled inflation; designed for adult home use; first on-sale date not establishedSemi-automatic oscillometric upper-arm measurement. User inflates cuff with supplied pump ball; electronic measurement follows during deflation. Quick-acting valve releases remaining cuff air. No automatic triple-measurement mode. First actual retail date unresolved
AccuracyMaker specification ±3 mmHg pressure and ±5% pulse; pressure range 0–299 mmHg, systolic 60–279 mmHg, diastolic 40–200 mmHg and pulse 40–180 beats/minExact April 2021 guide gives blood pressure 20–280 mmHg and pulse 40–200 bpm; global support instead lists systolic 60–255 and diastolic 40–200 mmHg. Asia-Pacific product page gives pulse upper limit 199 bpm. Static pressure accuracy ±3 mmHg and pulse accuracy ±5%; cuff-pressure display 0–299 mmHg is separate
Clinical validationA&D UK claims ESH 2010 validation; Medaval identifies the UK BIHS recommendation as manufacturer-declared equivalence to the UA-651 and says the equivalence study is unpublished. The US Validate BP listing covers UA-611 under ANSI/AAMI/ISO 81060-2:2018 with different cuff options; this does not establish direct validation of the supplied UK cuff configurationMaker guide claims shared measuring technology with BP 3BTO-H tested under ESH protocol and broader special-population suitability. Medaval reports no peer-reviewed exact-device validation evidence; bounded STRIDE search did not recover exact N2 Easy acceptance. Independent exact validation remains unresolved, so headline validation is omitted
FeaturesIrregular heartbeat indicator, pressure progress bar, WHO classification indicator, stored-reading average and automatic power-off; irregular heartbeat indication is not a specific AFib diagnosisGeneric IHB, termed PAD on Asia-Pacific product page; not AFib-specific screening or ECG. Hypertension warning, pump-up and air-discharge indicators, low-battery warning and automatic switch-off after approximately one minute. No date/time controls
Cuff
Cuff fits22–32 cm22–32 cm
CuffSupplied latex-free and metal-free SlimFit cuff fits 22–32 cm; optional large cuff fits 31–45 cm, sold separately; UK manual says this model is not designed for a small cuffAsia-Pacific bundle supplies conical M 22–32 cm cuff. April 2021 guide lists optional S 17–22 and L 32–42 cm; L cuff requires corresponding black deflation valve. Optional S cuff needs original cuff connector. Pump and cuff can use either of the two labelled pneumatic sockets; these are not dual-arm electronic measurement channels
Connections and memory
Users–1
Readings stored30 (better)1
Bluetooth and appNoNo
MemoryLast 30 blood pressure and pulse measurements, with stored-reading average; no separate user memories listedOne shared stored reading, replaced by the next result. Hold ON/OFF while switched off to recall. No multi-reading history, user selector, timestamps or averaging function in exact guide
AppNo Bluetooth; readings can be recorded manuallyNo Bluetooth, USB or PC-transfer function in exact controls
PowerFour AA 1.5 V batteries; maker claims approximately 700 measurements with LR6 alkaline batteries or 200 with R6P manganese batteries at 180 mmHg and 23°C. UA-611 has no DC socket; the optional adapter in the shared manual applies only to UA-651Two 1.5 V AA alkaline cells; approximately 1500 measurements with new batteries. NiMH cells require external charging. No mains-adaptor socket or operating instructions in exact labelled controls; generic adaptor wording in guarantee exclusions does not establish mains support
Design
Display detailsSystolic pressure, diastolic pressure and pulse shown with battery, irregular heartbeat and pressure/classification indicatorsLCD showing systolic, diastolic, pulse, IHB, stored-value, battery and inflation/deflation indicators. One ON/OFF button and two pneumatic sockets. IP20
SizeApproximately 96 × 68 × 130 mm (width × height × depth); approximately 230 g excluding batteries136 × 82 × 50 mm; 205 g including batteries in exact guide and global support
In the boxUA-611 monitor, 22–32 cm SlimFit cuff, instruction manual and four AA batteries in the UK packAsia-Pacific page lists monitor, two AA alkaline cells, pump ball, blood-pressure diary and M 22–32 cm cuff. Regional bundle scope retained; accessories are not inferred from other Microlife monitors